Instruction Manual CON 6/TDS 6
32
6.6 Normalization
Temperature
You can set the meter to normalize its conductivity measurements to a standard
temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1) Enter the advanced setup as described in
Section 6.1.
2) Press
▲
or
▼
key until [
t.nr °C
] appears on the
LCD. Press
ENTER
key.
3) Press
▲
or
▼
key to select either [
25.0 °C
] or
[
20.0 °C
].
4) Press
ENTER
key to select. The meter will take
you back to the menu, [
t.nr °C
].
5) Press
▲
or
▼
key to move to the next menu or
press
CAL
to exit to measurement mode.
6.7 Single-Point
Calibration
Single-point calibration refers to calibrating one conductivity value and uses it
for the entire 5 conductivity ranges.
By selecting [
no
] to single-point calibration, you can perform calibration for each
conductivity range.
1) Enter the advanced setup as described in
Section 6.1.
2) Press
▲
or
▼
key until [S.P.CA] appears on the
LCD. Press
ENTER
key.
3) Press
▲
or
▼
key to select either [
Yes
] or
no
].
4) Press
ENTER
key to select. The meter will take
you back to the menu, [
S.P.CA
].
5) Press
▲
or
▼
key to move to the next menu or
press
CAL
to exit to measurement mode.
C
°
C
°
C
°
HOLD
ENTER
HOLD
ENTER