Instruction Manual CON 6/TDS 6
29
6.2
Select Cell Constant
This meter lets you select a cell constant of K = 1.0, 10,
or 0.1.
Use a cell of K = 1.0 for midrange
measurements
Use a cell of K = 10 for high range
measurements (above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range
measurements (below 20 µS or 10 ppm).
The cell included with your meter has a cell constant of
K = 1.0.
1) Enter the advanced setup as described in
Section 6.1.
2) Press
▲
or
▼
key until [
CELC
] appears on the
LCD. Press
ENTER
key.
3) Press
▲
or
▼
key to select either “1.0”, “0.1” or
“10.0”. Ensure the cell constant selected
correspond with the conductivity electrode you are
using.
4) Press
ENTER
key to select. The meter will take
you back to the menu, [
CELC
].
5) Press
▲
or
▼
key to move to the next menu or
press
CAL
to exit to measurement mode.
HOLD
ENTER