Oakton CON 6 Скачать руководство пользователя страница 1

 

Technology Made Easy ...

Instruction Manual 

CON 6/TDS 6  

Hand-held Conductivity/TDS Meter 

 

 

 

68X243618 

ver. 0 11/02 

Содержание CON 6

Страница 1: ...Technology Made Easy Instruction Manual CON 6 TDS 6 Hand held Conductivity TDS Meter 68X243618 ver 0 11 02...

Страница 2: ...in the use of the Conductivity and TDS handheld meters please do not hesitate to contact the nearest Eutech Instruments Oakton Instruments Authorized Distributor Eutech Instruments Oakton Instruments...

Страница 3: ...ATION CONDUCTIVITY 16 4 5 MANUAL CALIBRATION CONDUCTIVITY OR TDS 18 4 6 TEMPERATURE CALIBRATION 19 5 MEASUREMENT 20 5 1 WITH AUTOMATIC TEMPERATURE COMPENSATION ATC 20 5 2 WITHOUT ATC MANUAL TEMPERATUR...

Страница 4: ...13 ADDENDUM 1 CALIBRATION TIPS 45 14 ADDENDUM 2 CALCULATING TDS CONVERSION FACTOR 46 15 ADDENDUM 3 CALCULATING TEMPERATURE COEFFICIENTS 47 16 WARRANTY 49 17 RETURN OF ITEMS 50...

Страница 5: ...ved Solids TDS and Temperature C This sturdy meter measures up to 5 different ranges with auto ranging capability that switches to appropriate measuring range automatically Your meter includes a condu...

Страница 6: ...Instruction Manual CON 6 TDS 6 2 mS ON OFF HOLD ENTER CAL MODE CON 6 Conductivity Meter C...

Страница 7: ...on is activated and LO low battery condition See Figure below pptm HO LO mS 1 2 3 4 5 6 7 C LCD and Customized Annunciators for CON 6 TDS 6 meter 1 Primary display 4 Temperature indicator 2 Parts Per...

Страница 8: ...NTER Press to confirm values in calibration mode and to confirm selections in SETUP mode In Calibration Mode Press to scroll through calibration values In Setup Mode Press to scroll through the setup...

Страница 9: ...meter into rubber boot slide in from the top of meter before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessa...

Страница 10: ...ry Replacement A LO annunciator in the LCD alerts you when battery power is running low See Figure below Replace with the same type as recommended by the manufacturer LO S LO Battery Condition Caution...

Страница 11: ...l resistant properties It provides fast temperature response and reduces air entrapment which makes it easy to obtain accurate stable readings The probe materials used which have good chemical durabil...

Страница 12: ...during measurement and calibration as it may affect your readings 2 We recommend that you do not submerge the probe above the protective guard You can submerge the cable for brief periods of time but...

Страница 13: ...nsert the mini phono jack of temperature sensor into the socket on the meter as shown below 4 Unplug the phono jack when not in use or when you want to measure Conductivity or TDS without any temperat...

Страница 14: ...OFF ppm pptmV pH MIN MAX LO HO F C mS pptm 3 6 Switching the Meter On When switching the meter on it will go through a series of display showing the various setup parameters For CON 6 Measurement Mod...

Страница 15: ...0 3 Third screen shows t 25 0 C which is the Normalization Temperature You can set Normalization Temperature at either 25 C or 20 C Refer to Section on Advance Setup Default value is 25 C 4 Fourth scr...

Страница 16: ...tivity TDS Temperature Measurement Mode To switch between Conductivity TDS measurement mode and Temperature measurement mode simply press the MODE key The customized annunciator helps indicate the mea...

Страница 17: ...0 00 10 00 ppm 3 00 to 8 50 ppm 0 0 200 0 S 60 0 to 170 0 S 10 0 100 0 ppm 30 0 to 85 0 ppm 0 2000 S 600 to 1700 S 100 1000 ppm 300 to 850 ppn 0 00 20 00 mS 6 00 to 17 00 mS 1 00 10 00 ppt 3 00 to 8 5...

Страница 18: ...uring in the mid ranges and you wash the probe in deionized water and store it dry calibrate the meter once a month If you take measurements at extreme temperatures calibrate at least once a week Ensu...

Страница 19: ...ome calibration procedure The known calibration standards used for automatic calibration are Meter Temperature Coefficient tc C Calibration Standards Range 25 C 1 84 S for 0 200 S cm 2 1413 S for 0 20...

Страница 20: ...ration standard Immerse the probe tip beyond the upper steel band see Figure in Section 3 4 Stir the probe gently to create a homogeneous sample Allow time for the reading to stabilize 4 Press CAL key...

Страница 21: ...solution is below 0 C or above 50 C 0 C t C 50 C an error message Err 2 will be displayed when performing the auto calibration and meter will return to measurement mode 3 All new calibration data wil...

Страница 22: ...upper steel band see Figure in Section 3 4 Stir the probe gently to create a homogeneous sample Allow time for the reading to stabilize 4 Press CAL key to enter conductivity calibration mode The CA in...

Страница 23: ...E key to select temperature measurement mode See Section 3 7 3 Press CAL key to start temperature calibration process 4 Dip the probe into a solution with known temperature for example a temperature b...

Страница 24: ...perature compensation 5 1 With Automatic Temperature Compensation ATC For ATC make sure the phono jack of the probe see Figure in Section 3 5 is securely inserted The conductivity TDS reading displaye...

Страница 25: ...asurement is disconnected from the meter See Figure in Section 3 5 2 Switch on the meter and if necessary press MODE key to select temperature measurement mode See Section 3 7 3 Press CAL key to start...

Страница 26: ...sample take care to ensure that the liquid level is above its upper steel band Stir the probe gently in the sample to create a homogenous sample See Figure in Section 3 4 5 4 Using Manual Ranging Fun...

Страница 27: ...mS cm Manual ranging 0 200 0 mS cm S S S NOTE If the value of the solution you are measuring is higher than the range selected Or will appear on the primary display Press RANGE until the correct rang...

Страница 28: ...nt mode 1 To hold a measurement press the HOLD key while in measurement mode HO will appear on the display 2 To release the held value press the HOLD again Continue to take measurements NOTE This mete...

Страница 29: ...ys for 2 seconds First release ON key first before releasing the MODE key 3 StUP indicator will appear momentarily and CELC will appear next 4 Overviews of CON 6 and TDS 6 s Setup Menu as follows Ente...

Страница 30: ...C or 25 C Default value is 25 C Adjust TDS factor from 0 4 to 1 0 Default value is 0 5 Available in TDS 6 meter only Select Single Point Calibration Choice of Yes or No Default value is Yes User re se...

Страница 31: ...r Off Press ON OFF and MODE keys simultaneously for 2 seconds release ON OFF key first then release MODE key a second later A StUP indicator will appear for 1 5 seconds before showing the first menu O...

Страница 32: ...r Off Press ON OFF and MODE keys simultaneously for 2 seconds release ON OFF key first then release MODE key a second later A StUP indicator will appear for 1 5 seconds before showing the first menu O...

Страница 33: ...or 10 ppm The cell included with your meter has a cell constant of K 1 0 1 Enter the advanced setup as described in Section 6 1 2 Press or key until CELC appears on the LCD Press ENTER key 3 Press or...

Страница 34: ...ill take you back to the menu ACAL 5 Press or key to move to the next menu or press CAL to exit to measurement mode 6 4 Setting the TDS Factor for TDS 6 The concentration of salts dissolved in solutio...

Страница 35: ...ductivity per degree of temperature it is expressed in percent per C Entering the exact temperature coefficient of your solution lets you accurately compensate temperature for almost any solution You...

Страница 36: ...y to move to the next menu or press CAL to exit to measurement mode 6 7 Single Point Calibration Single point calibration refers to calibrating one conductivity value and uses it for the entire 5 cond...

Страница 37: ...nce activated the meter s settings and calibration data will be erased and always exercise caution as meter reset is not reversible 1 Enter the advanced setup as described in Section 6 1 2 Press or ke...

Страница 38: ...e you soak your probe Do not immerse the probe in oily solutions Clean the electrode thoroughly by stirring it in a mild detergent bath or isopropyl alcohol Wipe the probe with a soft tissue paper Rin...

Страница 39: ...atteries with correct polarity c Replace batteries Unstable readings a Air bubbles in probe b Dirty probe c Probe not deep enough in sample d External noise pickup or induction caused by nearby electr...

Страница 40: ...calibration error Auto calibration is performed outside the temperature range 0 50 C Check the temperature and make sure that it is within the acceptable range If message persists return unit Err 3 C...

Страница 41: ...00 10 00 to 100 0 Up to 200 ppt depending on the TDS factor setting Resolution 0 01 0 1 1 ppm 0 01 0 1 ppt Accuracy 1 F S Temperature Range 10 0 to 110 0 C Resolution Accuracy 0 1 C 0 5 for C Cell Con...

Страница 42: ...HOLD Function Auto Power off 20 minutes after last key operation Inputs BNC for conductivity and phono jack for temperature Display Single Custom LCD Power Requirements 4 AAA Batteries Battery Life 1...

Страница 43: ...of k 1 0 EC CONSEN91B 35606 55 EC CON603k 35606 10 CON 6 portable TDS meter complete with conductivity probe of k 1 0 EC CONSEN91B 35606 55 EC TDS603K 35606 15 3 ring SS Ultem body Electrode with ATC...

Страница 44: ...libration Solution in 480 ml leak proof bottle 1 pint EC CON 2764BT 00653 20 10 S conductivity standard sachet 20 ml x 20 pcs EC CON 10BS 35653 09 447 S Conductivity Sachets 20 units x 20 ml per box E...

Страница 45: ...trical resistance of a solution C 1 R where C is conductance siemens R is resistance ohms One can combine Ohms law with the definition of conductance and the resulting relationship is C I E where I is...

Страница 46: ...ver this term has been replace by the term siemen Conductivity measurements are reported as Siemens cm since the value is measured between opposite faces of a cell of a known cubic configuration With...

Страница 47: ...particularly where high purity water is involved The unit most often used to describe resistivity is megohm cm which are simply the reciprocal of conductivity S cm The chart below shows the relations...

Страница 48: ...er degree depends on the composition of the solution being measured However for most medium range salt concentration in water 2 per degree works well Extremely pure water exhibits a temperature coeffi...

Страница 49: ...t specified 1 F S accuracy If you are measuring in the mid ranges and you washed the probe in deionized water and stored it dry calibrate the meter at least once a month Wet the probe for 10 minutes b...

Страница 50: ...ns Actual TDS Value from the solution bottle label or as a standard you make using high purity water and precisely weighed salts Actual Conductivity Value measured using a properly calibrated Conducti...

Страница 51: ...e 1 Immerse the probe into a sample of your solution and adjust the temperature coefficient to 0 that is no compensation by following instructions as described in Section 6 5 Temperature Coefficient 2...

Страница 52: ...coefficient of your solution according to the formula shown above 5 Enter the temperature coefficient you calculated into the meter Refer to Section 6 5 Temperature Coefficient The calculated temperat...

Страница 53: ...within the designated period please return freight pre paid and correction will be made without charge Eutech Instruments Oakton Instruments will determine if the product problem is due to deviations...

Страница 54: ...he reason the items are to be returned For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Eutech Instruments Oakton Instrument...

Страница 55: ...NOTES...

Страница 56: ...te listed below Oakton Instruments P O Box 5136 Vernon Hills IL60061 USA Fax 1 847 247 2984 www 4oakton com www oaktoninstruments com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Crescent 04 16 24 Sin...

Отзывы: