Instruction Manual
CON 11/110
70
14.2 CON 110 Factory Default Setting
PROGRAM FUNCTION
PARAMETER
OPTING KEYS
PARAMETER
OPTIONS
DEFAULT
SETTING
P 1.0
Unit Configuration
Parameter Title
Only
Parameter Title
Only
P 1.1
Selection of READY Indicator and Auto
HOLD Function
MI/
▲
; MR/
▼
ON ; OFF,
HOLD
ON
P 1.2
Selection of °C and °F indicator
MI/
▲
; MR/
▼
°C, °F
°C
P 1.3
Selection of Automatic or Manual
Temperature Compensation
MI/
▲
; MR/
▼
YES,
nO YES
P 1.4
Selection of TDS Conversion Factor
MI/
▲
; MR/
▼
0.40 to 1.00
0.50
P 2.0
Viewing previous calibration data
Parameter Title
Only
Parameter Title
Only
P 2.1
First range calibration point
Viewing Only
Viewing Only
_ _ _
P 2.2
Second range calibration point
Viewing Only
Viewing Only
_ _ _
P 2.3
Third range calibration point
Viewing Only
Viewing Only
_ _ _
P 2.4
Fourth range calibration point
Viewing Only
Viewing Only
_ _ _
P 2.5
Fifth range calibration point
Viewing Only
Viewing Only
_ _ _
P3.0
Viewing electrode data
Parameter Title
Only
Parameter Title
Only
P3.1
Effective cell constant for first range
Viewing Only
Viewing Only
K = “1.00”
P3.2
Effective cell constant for second range
Viewing Only
Viewing Only
K = “1.00”
P3.3
Effective cell constant for third range
Viewing Only
Viewing Only
K = “1.00”
P3.4
Effective cell constant for fourth range
Viewing Only
Viewing Only
K = “1.00”
P3.5
Effective cell constant for fifth range
Viewing Only
Viewing Only
K = “1.00”
P4.0:
Auto Power OFF
MI/
▲
; MR/
▼
YES,
nO YES
P5.0
Reset to factory defaults
Parameter Title
Only
Parameter Title
Only
P5.1
Resetting of all calibration and clear all
calibration data
MI/
▲
; MR/
▼
nO,
YES nO
P5.2
Resetting of all User settings to factory
defaults condition
MI/
▲
; MR/
▼
nO,
YES nO
P6.0 Memory
clear
MI/
▲
; MR/
▼
nO,
YES nO
P7.0 Temperature
Parameter Title
Only
Parameter Title
Only
P7.1
Setting of temperature coefficient
MI/
▲
; MR/
▼
0.0 to 10.0
2.1
P7.2
Setting of normalisation temperature
MI/
▲
; MR/
▼
20.0,
25.0 25.0
P8.0
Selection of cell constant
MI/
▲
; MR/
▼
1.0, 10.0, 0.1
K = “1.0”
P9.0 Automatic
Single Point Calibration
Parameter Title
Only
Parameter Title
Only
P9.1
Selection of Automatic or Manual
Calibration
MI/
▲
; MR/
▼
YES,
nO YES
P9.2
Selection of Single or Multi Point
Calibration
MI/
▲
; MR/
▼
YES,
nO YES
NOTE: The HOLD/ENTER key is used to confirm every parameter opted. The CAL/MEAS on the other hand will
escape the set up operation to return to the measurement mode.