manualshive.com logo in svg
background image

 

37003-02 HumidityLogR Set-Up 

The interval may be set to any time from once every 4 seconds to once 
every hour. 
This example will set the thermohygrometer to print live readings once 
every 4 seconds.  
Press 

SETUP

, then 

PRINT

. The display indicates “minutes:seconds” with the 

last printing interval setup shown. The seconds will be flashing. 
 

1.

 

Press 

MAX 

(up) or 

MIN 

(down), until the flashing digits indicate 

“04” seconds. 

2.

 

Press 

HOLD 

(enter). The display will show the minutes flashing. 

3.

 

Press 

HOLD 

again. The large upper display will show “HP”, “300”, 

“600”, “1200” or “2400”. 

4.

 

Press 

MAX 

until “2400” shows in the upper display. 

5.

 

Press 

SETUP 

to exit. 

Aim the top of the unit at the red window of the adapter at a distance of 12 
inches or less. 
 

Transferring Live Readings 

To begin data transfer, press 

PRINT

. Whenever the “PRINT” annunciator is 

displayed on the unit, and the red light inside the RS-232 adapter is blinking, 
data should appear on your computer screen. 
 

Transferring Stored Readings 

The “STO” annunciator on the unit’s display indicates that readings are 
stored in memory. Press RECALL then PRINT to begin data transfer. “RCL” 
and “PRINT” annunciators will be on the thermometer display, and the 
number of readings remaining to be transferred will count down. The red 
light inside the RS232 adapter should be blinking. 
 

WINDOWS XP Hyperterminal Set-up 

1. Click on Accessories, select Communications, then Hyperterminal. 
2. Under NAME: Enter a name for your file and click on 

OK

3. Under CONNECT USING: Choose COM1. 

3. Click on 

OK

4. Under COM1 PROPERTIES: Enter 2400, hit Tab, 8, hit Tab, None, hit Tab, 

and 1.  

5. Click on 

OK

.  

6. Check that the red light inside the window of the adapter is on. 
7. Align the infrared windows and select PRIN.  

 
 
 
 

5

 

Содержание 91100-85

Страница 1: ...RS 232C Adapter Model 91100 85 USER GUIDE 68X343601 Rev 1 05 07...

Страница 2: ...1 Connecting Your RS232C Adapter 2 Setting Up Your Devices 3 Setting Up Your Thermo Devices 4 91100 50 DuaLogR Set Up 4 93210 50 and 93410 50 ThermoLogR Set Up 4 37003 02 HumidityLogR Set Up 5 Transfe...

Страница 3: ...PCD 650 PC 650 PD 650 CD 650 pH 620 610 600 COND 610 600 DO 600 What this adapter enables you to do The RS232 Adapter receives your data via Infrared signals and lets you print directly through a PC o...

Страница 4: ...ure that your PC is plugged into a power source or has sufficient battery power 3 If using the microprinter plug in the power supply adapter to a power source and connect the cable to the microprinter...

Страница 5: ...PRINT MODE and change IrDA to LED by using the c or darrows 7 Then hit ENTER to move one step down to DATA FORMAT 8 If you re printing on a PC with the Hyperterminal software or on the microprinter c...

Страница 6: ...P 300 or 600 5 Press MAX until 600 shows use 300 for HYPERTERMINAL in the upper display 6 Press SETUP to exit Aim the top of the unit at the red window of the adapter at a distance of 12 inches or les...

Страница 7: ...egin data transfer press PRINT Whenever the PRINT annunciator is displayed on the unit and the red light inside the RS 232 adapter is blinking data should appear on your computer screen Transferring S...

Страница 8: ...uaLogR see below When all of the data has transferred click on Transfer Capture Text then Stop The next time you run HYPERTERMINAL you may select DuaLogR ht The HYPERTERMINAL program will automaticall...

Страница 9: ...roprinter Communication ECMICROPRNTR01 01X230001 Serial Impact Microprinter 110 VAC ECMICROPRNTR02 01X230002 Serial Impact Microprinter 220 VAC ECMICROPRNTRO2CLK 01X230004 Serial Impact Microprinter w...

Отзывы: