manualshive.com logo in svg
background image

13

12

6. 

Press ENTER to confirm calibration

The meter is now calibrated to the 
current buffer.

See figure 

• If you are performing multipoint 

calibration, go to step 7. 

• If you are performing one-point 

calibration, go to step 9.

7. 

Rinse the probe with de-ionized water
or a rinse solution, and place it in the
next pH buffer. 

8. 

Follow steps 5 to 8 for additional 
calibration points. 

9. 

When calibration is complete, 
press CAL/MEAS to return to pH 
measurement mode

.

Notes

To exit from pH Calibration mode without
confirming calibration,

DO NOT press

ENTER 

in step 6. Press CAL/MEAS instead.

If the selected buffer value is not within
±1.0 pH from the measured pH value: the
electrode and buffer icon blink and the 
ERR annunciator appears in the lower left
corner of the display.

To limit the number of pH buffer values
available during calibration, see section
P3.2 on page 36.

pH

READY

CAL

pH

D

D

pH Calibration

1. 

If necessary, press the MODE key to
select pH mode

. The pH indicator 

appears in the upper right hand corner 
of the display.

See figure 

2. 

Rinse the pH electrode and DO probe
thoroughly with de-ionized water or a
rinse solution.

Do not wipe the probe;

this causes a build-up of electrostatic
charge on the glass surface.

3. 

Dip both probes into the calibration
buffer. 

The end of both probes must be

completely immersed into the sample. Stir
the probe gently to create a homogeneous
sample.

4. 

Press CAL/MEAS

to enter pH calibration

mode. The CAL indicator lights. The 
primary display will show the measured
reading while the smaller secondary 
display will indicate the pH standard
buffer solution. 

See figure 

5. 

Wait for the measured pH value to 
stabilize.

See figure

pH

READY

CAL

pH

C

C

B

pH

CAL

pH

B

pH

ATC

MEAS

 C

A

A

pH calibration

NOTE: 

We recommend that you perform at least a 2-point calibration using standard
buffers that bracket (one above and one below) the expected sample range. 
You can perform a 1-point calibration, but make sure that the buffer value is
close to the sample value you are measuring.

The temperature sensing element is in the DO probe. For temperature 
compensated readings and calibration, immerse the DO probe into the 
calibration buffer while calibrating pH.

If the READY indicator has been activated 
(set up program P3.1—see page 35), the READY
annunciator lights when the reading is stable.

4.3

Содержание 35632-Series

Страница 1: ...s Portable Waterproof pH Dissolved Oxygen Meter pH Dissolved Oxygen C F Meter pH DO 300 Series WATERPROOF MODE HOLD ENTER ON OFF CAL MEAS SETUP ATC READY MEAS C IP 67 Notes Write down the name and inf...

Страница 2: ...efault settings 38 8 6 saturation mode Program P1 0 Dissolved oxygen parameters Pressure adjustment mode 39 8 7 saturation mode Program P2 0 Offset adjustment 40 8 8 saturation mode Program P3 0 Previ...

Страница 3: ...lay shows the temperature of the reading The display also shows error messages keypad functions and program functions ATC READY MEAS CAL HOLD ERR F pH SETUP ON OFF C ppm mg l ppt 2 1 Primary display 5...

Страница 4: ...pm calibration mode NOTE Temperature calibration is available from both pH and DO mg l ppm calibration modes see page 19 for directions In advanced set up mode Press CAL MEAS to return to main menu fr...

Страница 5: ...connector Pull probe away from the meter CAUTION Do not pull on the probe cord or the probe wires might disconnect DO probe with 6 pin connector locking ring pH Electrode with BNC connector top view o...

Страница 6: ...libration at pH 1 68 4 01 7 00 10 01 and 12 45 to ensure accuracy across the entire pH range of the meter The meter automatically recognizes and calibrates to these standard buffer values which makes...

Страница 7: ...de and DO probe thoroughly with de ionized water or a rinse solution Do not wipe the probe this causes a build up of electrostatic charge on the glass surface 3 Dip both probes into the calibration bu...

Страница 8: ...and returns to Measurement mode See figure ATC READY MEAS C READY CAL ATC READY MEAS C 4 5 A A B C B C Notes The reading in the primary display in step 3 must read at 50 or above for the calibration...

Страница 9: ...reading in the primary display in step 3 must read at 10 or under for the calibra tion to work correctly Whenever an error occurs during calibration the ERR indicator appears in the lower left hand c...

Страница 10: ...l ppm Measurement mode 2 Press the CAL MEAS key to enter pH or mg l ppm calibration mode The CAL indicator will appear above the primary display See figure 3 While in pH or mg l ppm calibration mode p...

Страница 11: ...the READY indicator on or off in the SETUP program P5 1 for saturation measurements page 43 or program P4 1 page 51 for mg l ppm measurements Taking measurements with the auto endpoint feature selecte...

Страница 12: ...of the LCD The ATC indicator appears in the lower right hand corner to indicate Automatic Temperature Compensation If the ATC indicator does not appear make sure DO probe is plugged into the meter Se...

Страница 13: ...9 1 Switch the meter on Press the MODE key to select pH mode 2 Press the CAL MEAS key to enter pH calibration mode The CAL indicator will appear above the primary display 3 While in pH calibration mod...

Страница 14: ...mmest light to level 8 brightest light 6 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CAL MEAS key to return to measurement mode SETUP SETUP A A B B A This fea...

Страница 15: ...E Probe data COF Configuration rSt Reset meter to factory default CAL Calibration data dPr Dissolved oxygen parameters pH Saturation mg l ppm LCd Backlit display rSt Reset meter to factory default Fro...

Страница 16: ...ctions on page 45 SETUP rSt Reset to factory default Reset to factory default settings Instructions on page 46 SETUP From the mg l ppm measurement mode dPr Dissolved oxygen parameters Select Hg or Pa...

Страница 17: ...recall previous calibration data which helps you know when to recalibrate your meter This is a view only mode 8 2 pH SETUP A A pH SETUP B pH SETUP C B C P2 1 Electrode offset From measurement mode 1...

Страница 18: ...f ON switches the READY indicator on ON and HOLD together switches the auto endpoint feature on 6 Press the ENTER key to confirm selec tion and to proceed to step 4 of P3 2 Press the CAL MEAS key to r...

Страница 19: ...ration mode after you have calibrated to your selected number of points P3 3 Selecting C or F This meter lets you select between C and F units for temperature readings From measurement mode 1 Press th...

Страница 20: ...meters This sub group lets you adjust the barometric pressure Pressure adjustment mode Barometric pressure is vital to correct dissolved oxygen measurements You need to enter the correct barometric pr...

Страница 21: ...ant to return to measure ment mode press CAL MEAS SETUP READY MEAS SETUP A A B B saturation mode P3 0 Previous calibration information This sub group shows you the previous calibration data along with...

Страница 22: ...ey to enter SETUP mode 3 Press the and keys to scroll through subgroups until you view parameter P5 0 in the lower display See figure 4 Press the ENTER key to select parameter P5 1 rdY Ready See figur...

Страница 23: ...e brightness of the backlit LCD Selecting a dimmer backlighting level helps conserve batteries 8 11 From measurement mode 1 Press the MODE key to select the saturation measurement mode 2 Press SETUP k...

Страница 24: ...barometric pressure and salinity P1 1 Pressure adjustment mode Barometric pressure is vital to correct dissolved oxygen measurements You need to enter the correct barometric pressure of the area you...

Страница 25: ...P1 0 If you want to return to measure ment mode press CAL MEAS SETUP ppt SETUP D D E E mg l ppm mode P2 0 Previous calibration information This sub group shows you the previous calibration data along...

Страница 26: ...function The first program lets you select READY indicator on to indicate when the reading is stable READY indicator off for faster meter response Auto endpoint function on Select auto endpoint on to...

Страница 27: ...Selecting C or F temperature readout This meter lets you select between C and F units for temperature readings From measurement mode 1 Press the MODE key to select mg l ppm mode 2 Press SETUP key to e...

Страница 28: ...rogram lets you reset all parameters to factory default settings This clears all calibration data memory and any other setup functions you might have changed The LCD contrast setting will be reset to...

Страница 29: ...solution to store your electrode see page 64 65 for ordering information Also you can store in a pH 4 buffer with 1 100 part of saturated KCl Other pH buffers are OK for storage but NEVER use distill...

Страница 30: ...ert notches on installation tool into slots on membrane lock Then unscrew membrane lock from membrane cap B C B C 5 Using the installation tool unscrew and remove the membrane lock from the membrane c...

Страница 31: ...ies not in place a Check that batteries are in no display place and making good contact b Batteries not in correct b Reinsert batteries with polarity and correct polarity c Weak batteries c Replace ba...

Страница 32: ...00 0 0 0 to 50 0 C 20 00 mg l ppm 32 0 to 122 0 F Resolution 0 01 pH 0 01 mg l ppm 0 1 0 1 C 0 1 F Accuracy 0 01 pH 1 5 full scale 1 5 full scale 0 3 C 0 5 F Calibration up to 5 points one or two poin...

Страница 33: ...dy have 35640 74 Replacement DO membranes 5 pack 35640 75 Replacement DO membranes 25 pack 35640 79 DO membrane installation tool Required for membrane installation Not required if you purchase replac...

Страница 34: ...sate for these factors This meter automatically compensates for tempera ture readings It also allows you to enter a salinity correction factor and the baromet ric pressure to correct for this variabil...

Страница 35: ...authorized returns NOTE We reserve the right to make improvements in design construction and appearance of products without notice 16 Appendix 3 Factory Default Settings Resetting the meter to factory...

Отзывы: