DO Membrane Replacement

Replacement of the membrane is required only
when you cannot calibrate the probe, or if the
membrane is damaged.

Typical membrane damages are punctures 
or wrinkles caused during measurements or
cleaning.

To order replacement probe components or a
replacement probe, see the “Accessories” section
on page 64.

To replace the probe membrane:

1.

Pull off the probe guard.

2.

Unscrew the membrane cap from the probe.

3.

Hold the probe under a water tap and brush
away the white oxide on the cylindrical anode
with a stiff plastic brush—do not use metal
cleaning material.

4. 

It the cathode has any deposits, remove them
with a light scouring powder. Do not polish 
the cathode. 

NOTE: 

If you have purchased a replacement 
membrane module with preinstalled
membrane, skip to step 12.

CONTINUED ON NEXT PAGE

Temperature

sensor

Anode

Membrane

lock

Membrane

O-ring

Membrane 

cap

Probe 
guard

Cathode

9.3

A

Installation tool

Carefully place
O-ring, then
membrane,
into membrane
cap.

Insert notches
on installation
tool into slots on
membrane lock.
Then, unscrew 
membrane 
lock from 
membrane cap.

B

C

B

C

5.

Using the installation tool, unscrew and
remove the membrane lock from the 
membrane cap.

See figure

6.

Remove the membrane and O-ring. 
Discard both.

7.

Rinse the membrane cap and membrane lock
in tap water.

8.

Install a new O-ring inside the 
membrane cap.

9.

Install a new membrane. Make sure the 
membrane covers the O-ring all around its
circumference.

See figure

10.

Using the installation tool, screw the 
membrane lock back into the cap. Tighten
the lock firmly over the membrane and 
O-ring, but do not overtighten.

11.

Inspect the membrane for wrinkles. 
If wrinkles exist, remove the membrane 
and repeat steps 8-11.

12.

Fill the membrane cap with water and
inspect the bottom for leaks. If water drops
are leaking from the membrane, re-seat the
membrane on the O-ring (repeat steps 8-11).

13.

If the assembly is leak-free, empty the water
and fill the membrane cap with electrolyte
to the brim.

14.

Screw the cap onto the probe. 
Excess electrolyte will drain out.

15.

Replace probe guard.

16.

Calibrate the probe (see section 4.4-4.6 on
pages 14-18) after the % saturation readings
have stabilized. 

Содержание 35632-Series

Страница 1: ...s Portable Waterproof pH Dissolved Oxygen Meter pH Dissolved Oxygen C F Meter pH DO 300 Series WATERPROOF MODE HOLD ENTER ON OFF CAL MEAS SETUP ATC READY MEAS C IP 67 Notes Write down the name and inf...

Страница 2: ...efault settings 38 8 6 saturation mode Program P1 0 Dissolved oxygen parameters Pressure adjustment mode 39 8 7 saturation mode Program P2 0 Offset adjustment 40 8 8 saturation mode Program P3 0 Previ...

Страница 3: ...lay shows the temperature of the reading The display also shows error messages keypad functions and program functions ATC READY MEAS CAL HOLD ERR F pH SETUP ON OFF C ppm mg l ppt 2 1 Primary display 5...

Страница 4: ...pm calibration mode NOTE Temperature calibration is available from both pH and DO mg l ppm calibration modes see page 19 for directions In advanced set up mode Press CAL MEAS to return to main menu fr...

Страница 5: ...connector Pull probe away from the meter CAUTION Do not pull on the probe cord or the probe wires might disconnect DO probe with 6 pin connector locking ring pH Electrode with BNC connector top view o...

Страница 6: ...libration at pH 1 68 4 01 7 00 10 01 and 12 45 to ensure accuracy across the entire pH range of the meter The meter automatically recognizes and calibrates to these standard buffer values which makes...

Страница 7: ...de and DO probe thoroughly with de ionized water or a rinse solution Do not wipe the probe this causes a build up of electrostatic charge on the glass surface 3 Dip both probes into the calibration bu...

Страница 8: ...and returns to Measurement mode See figure ATC READY MEAS C READY CAL ATC READY MEAS C 4 5 A A B C B C Notes The reading in the primary display in step 3 must read at 50 or above for the calibration...

Страница 9: ...reading in the primary display in step 3 must read at 10 or under for the calibra tion to work correctly Whenever an error occurs during calibration the ERR indicator appears in the lower left hand c...

Страница 10: ...l ppm Measurement mode 2 Press the CAL MEAS key to enter pH or mg l ppm calibration mode The CAL indicator will appear above the primary display See figure 3 While in pH or mg l ppm calibration mode p...

Страница 11: ...the READY indicator on or off in the SETUP program P5 1 for saturation measurements page 43 or program P4 1 page 51 for mg l ppm measurements Taking measurements with the auto endpoint feature selecte...

Страница 12: ...of the LCD The ATC indicator appears in the lower right hand corner to indicate Automatic Temperature Compensation If the ATC indicator does not appear make sure DO probe is plugged into the meter Se...

Страница 13: ...9 1 Switch the meter on Press the MODE key to select pH mode 2 Press the CAL MEAS key to enter pH calibration mode The CAL indicator will appear above the primary display 3 While in pH calibration mod...

Страница 14: ...mmest light to level 8 brightest light 6 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CAL MEAS key to return to measurement mode SETUP SETUP A A B B A This fea...

Страница 15: ...E Probe data COF Configuration rSt Reset meter to factory default CAL Calibration data dPr Dissolved oxygen parameters pH Saturation mg l ppm LCd Backlit display rSt Reset meter to factory default Fro...

Страница 16: ...ctions on page 45 SETUP rSt Reset to factory default Reset to factory default settings Instructions on page 46 SETUP From the mg l ppm measurement mode dPr Dissolved oxygen parameters Select Hg or Pa...

Страница 17: ...recall previous calibration data which helps you know when to recalibrate your meter This is a view only mode 8 2 pH SETUP A A pH SETUP B pH SETUP C B C P2 1 Electrode offset From measurement mode 1...

Страница 18: ...f ON switches the READY indicator on ON and HOLD together switches the auto endpoint feature on 6 Press the ENTER key to confirm selec tion and to proceed to step 4 of P3 2 Press the CAL MEAS key to r...

Страница 19: ...ration mode after you have calibrated to your selected number of points P3 3 Selecting C or F This meter lets you select between C and F units for temperature readings From measurement mode 1 Press th...

Страница 20: ...meters This sub group lets you adjust the barometric pressure Pressure adjustment mode Barometric pressure is vital to correct dissolved oxygen measurements You need to enter the correct barometric pr...

Страница 21: ...ant to return to measure ment mode press CAL MEAS SETUP READY MEAS SETUP A A B B saturation mode P3 0 Previous calibration information This sub group shows you the previous calibration data along with...

Страница 22: ...ey to enter SETUP mode 3 Press the and keys to scroll through subgroups until you view parameter P5 0 in the lower display See figure 4 Press the ENTER key to select parameter P5 1 rdY Ready See figur...

Страница 23: ...e brightness of the backlit LCD Selecting a dimmer backlighting level helps conserve batteries 8 11 From measurement mode 1 Press the MODE key to select the saturation measurement mode 2 Press SETUP k...

Страница 24: ...barometric pressure and salinity P1 1 Pressure adjustment mode Barometric pressure is vital to correct dissolved oxygen measurements You need to enter the correct barometric pressure of the area you...

Страница 25: ...P1 0 If you want to return to measure ment mode press CAL MEAS SETUP ppt SETUP D D E E mg l ppm mode P2 0 Previous calibration information This sub group shows you the previous calibration data along...

Страница 26: ...function The first program lets you select READY indicator on to indicate when the reading is stable READY indicator off for faster meter response Auto endpoint function on Select auto endpoint on to...

Страница 27: ...Selecting C or F temperature readout This meter lets you select between C and F units for temperature readings From measurement mode 1 Press the MODE key to select mg l ppm mode 2 Press SETUP key to e...

Страница 28: ...rogram lets you reset all parameters to factory default settings This clears all calibration data memory and any other setup functions you might have changed The LCD contrast setting will be reset to...

Страница 29: ...solution to store your electrode see page 64 65 for ordering information Also you can store in a pH 4 buffer with 1 100 part of saturated KCl Other pH buffers are OK for storage but NEVER use distill...

Страница 30: ...ert notches on installation tool into slots on membrane lock Then unscrew membrane lock from membrane cap B C B C 5 Using the installation tool unscrew and remove the membrane lock from the membrane c...

Страница 31: ...ies not in place a Check that batteries are in no display place and making good contact b Batteries not in correct b Reinsert batteries with polarity and correct polarity c Weak batteries c Replace ba...

Страница 32: ...00 0 0 0 to 50 0 C 20 00 mg l ppm 32 0 to 122 0 F Resolution 0 01 pH 0 01 mg l ppm 0 1 0 1 C 0 1 F Accuracy 0 01 pH 1 5 full scale 1 5 full scale 0 3 C 0 5 F Calibration up to 5 points one or two poin...

Страница 33: ...dy have 35640 74 Replacement DO membranes 5 pack 35640 75 Replacement DO membranes 25 pack 35640 79 DO membrane installation tool Required for membrane installation Not required if you purchase replac...

Страница 34: ...sate for these factors This meter automatically compensates for tempera ture readings It also allows you to enter a salinity correction factor and the baromet ric pressure to correct for this variabil...

Страница 35: ...authorized returns NOTE We reserve the right to make improvements in design construction and appearance of products without notice 16 Appendix 3 Factory Default Settings Resetting the meter to factory...

Отзывы: