Demand Valve and Demand Valve Resuscitator - User Manual

4.  Servicing

4.1.   Routine Maintenance 

   

WARNING

   

The O-Two Demand Valve and Demand Valve Resuscitator 

are designed to provide respiratory support in all emergency 

situations. Failure to follow the maintenance and inspection 

routine could result in incorrect operation of the resuscitator.

To ensure proper operation of the Demand Valve and Demand 

Valve Resuscitator, regular inspection and checking of the demand 

valve and accessories for correct function should be undertaken 

by a responsible member of staff  on a regular basis, and it is 

recommended that a service record be maintained for each unit.

This check is to ensure that all of the accessories and demand 

valve components are present, the oxygen cylinder is full and that 

the demand valve is in working order.

MAINTENANCE

Resuscitation parameters should be checked at least every six 

months, and more frequently in high use applications. Units 

with test parameters outside of the ranges listed in the product 

specifi cations should not be used. O-Two Demand Valve are not 

designed for fi eld disassembly or service outside that indicated 

in this manual. Any malfunctioning units should be returned to 

the manufacturer or an Authorized O-Two Medical Technologies 

Dealer. Unauthorized repairs will nullify the product warranty.

Содержание 01DV2000

Страница 1: ...17MP3136 Rev 13 FEB 2020 USER MANUAL DEMAND VALVE 01DV3000 DEMAND VALVE RESUSCITATOR 01DV2000...

Страница 2: ...4 1 4 Performance Specification 5 2 Gas Supply 6 2 1 Gas Supply Connections 6 3 Operating Procedure 6 3 1 Demand Breathing 6 3 2 Manual Resuscitation 01DV2000 only 6 3 3 Operating Instructions 7 4 Se...

Страница 3: ...anty WARRANTY O Two equipment is manufactured from the finest quality materials Each individual part is subject to strict quality control tests to ensure exceptionally high standards The manufacturer...

Страница 4: ...IS IN USE DO NOT SMOKE OR USE NEAR OPEN FLAME EITHER DURING USE OR WHEN CHANGING THE CYLINDER 3 WHEN NOT IN USE ALWAYS TURN OFF THE CYLINDER 4 NEVER ALLOW OIL OR GREASE TO COME INTO CONTACT WITH ANY...

Страница 5: ...x Pressure Relief Valve 01DV2000 only 60 cmH2 O Audible Over Pressure Alarm 01DV2000 only YES Triggering Pressure 0 to 3 5 cmH2 O Operating Temperature 180 C to 500 C 00 F to 1220 F Storage Temperatur...

Страница 6: ...ow to the patient at a rate and volume in line with their inspiratory effort 3 2 Manual Resuscitation 01DV2000 only The O Two Demand Valve Resuscitator 01DV2000 has been fitted with a Manual Button t...

Страница 7: ...ing both the nose and mouth and hold the mask on his her face with a light pressure The sound of gas owing should be heard with patient s inspiratory efforts 6 Should manual ventilation is required fo...

Страница 8: ...a regular basis and it is recommended that a service record be maintained for each unit This check is to ensure that all of the accessories and demand valve components are present the oxygen cylinder...

Страница 9: ...should be undertaken after each patient use to maintain the equipment in a clean condition WARNING Do not immerse the demand valve into cleaning solution or disinfectant Discontinue to use and discar...

Страница 10: ...be required 4 Unscrew patient valve A from the unit body being careful to ensure that the diaphragm B is retained Fig1 5 Remove the diaphragm and anti entrainment ring C from the patient valve swivel...

Страница 11: ...s System 01FV4303 AFNR O Two 6 Foot 1 85 Meter O2 Supply Hose with AFNOR probe and 9 16 DISS Nut Ventilator Connection 01FV4303 AGA O Two 6 Foot 1 85 Meter O2 Supply Hose with AGA probe and 9 16 DISS...

Страница 12: ...tor In North America call Toll Free 1 800 387 3405 MedNet EC REP GmbH Borkstrasse 10 48163 M nster Germany EC REP SERIAL N0 www otwo com 45A Armthorpe Road Brampton ON Canada L6T 5M4 Telephone 1 905 7...

Отзывы: