Demand Valve and Demand Valve Resuscitator - User Manual

Cleaning / Disinfection Procedure

1.  Operate the manual button to blow out any contaminant from 

the patient valve (01DV2000 only).

2.  Ensure that the cylinder is turned off  (completely clockwise) 

and bleed any remaining gas out of the system by depressing 

the manual ventilation button (01DV2000 only).

3.  Remove the facemask from the Demand Valve (Resuscitator), 

and then disconnect the unit from the regulator. This is 

accomplished by counter-clockwise turning of the connection 

nut at the regulator end of the hose. A wrench may be required.

4.  Unscrew patient valve (A) from the unit body being careful to 

ensure that the diaphragm (B) is retained (Fig1). 

5.  Remove the diaphragm and anti-entrainment ring (C) from 

the patient valve swivel housing and wash these components 

thoroughly in a mild soap solution;

6.  Disinfect the face mask, patient valve and diaphragm using 

commonly available cold chemical disinfecting solutions. 

Rinse thoroughly with water after disinfection.

7.  Sterilization, if required, may be achieved by other methods 

compatible with the materials of manufacture of these 

components including Ethylene Oxide Gas and Cidex. 

8.  The demand valve body (D) and gas supply hose can be wiped 

over with a soft cloth and mild soap solution.

9.  Dry all components thoroughly.

10.  Re-assemble and check performance.

Содержание 01DV2000

Страница 1: ...17MP3136 Rev 13 FEB 2020 USER MANUAL DEMAND VALVE 01DV3000 DEMAND VALVE RESUSCITATOR 01DV2000...

Страница 2: ...4 1 4 Performance Specification 5 2 Gas Supply 6 2 1 Gas Supply Connections 6 3 Operating Procedure 6 3 1 Demand Breathing 6 3 2 Manual Resuscitation 01DV2000 only 6 3 3 Operating Instructions 7 4 Se...

Страница 3: ...anty WARRANTY O Two equipment is manufactured from the finest quality materials Each individual part is subject to strict quality control tests to ensure exceptionally high standards The manufacturer...

Страница 4: ...IS IN USE DO NOT SMOKE OR USE NEAR OPEN FLAME EITHER DURING USE OR WHEN CHANGING THE CYLINDER 3 WHEN NOT IN USE ALWAYS TURN OFF THE CYLINDER 4 NEVER ALLOW OIL OR GREASE TO COME INTO CONTACT WITH ANY...

Страница 5: ...x Pressure Relief Valve 01DV2000 only 60 cmH2 O Audible Over Pressure Alarm 01DV2000 only YES Triggering Pressure 0 to 3 5 cmH2 O Operating Temperature 180 C to 500 C 00 F to 1220 F Storage Temperatur...

Страница 6: ...ow to the patient at a rate and volume in line with their inspiratory effort 3 2 Manual Resuscitation 01DV2000 only The O Two Demand Valve Resuscitator 01DV2000 has been fitted with a Manual Button t...

Страница 7: ...ing both the nose and mouth and hold the mask on his her face with a light pressure The sound of gas owing should be heard with patient s inspiratory efforts 6 Should manual ventilation is required fo...

Страница 8: ...a regular basis and it is recommended that a service record be maintained for each unit This check is to ensure that all of the accessories and demand valve components are present the oxygen cylinder...

Страница 9: ...should be undertaken after each patient use to maintain the equipment in a clean condition WARNING Do not immerse the demand valve into cleaning solution or disinfectant Discontinue to use and discar...

Страница 10: ...be required 4 Unscrew patient valve A from the unit body being careful to ensure that the diaphragm B is retained Fig1 5 Remove the diaphragm and anti entrainment ring C from the patient valve swivel...

Страница 11: ...s System 01FV4303 AFNR O Two 6 Foot 1 85 Meter O2 Supply Hose with AFNOR probe and 9 16 DISS Nut Ventilator Connection 01FV4303 AGA O Two 6 Foot 1 85 Meter O2 Supply Hose with AGA probe and 9 16 DISS...

Страница 12: ...tor In North America call Toll Free 1 800 387 3405 MedNet EC REP GmbH Borkstrasse 10 48163 M nster Germany EC REP SERIAL N0 www otwo com 45A Armthorpe Road Brampton ON Canada L6T 5M4 Telephone 1 905 7...

Отзывы: