1.6. Performance Specifi cations
12. Do not disassemble any part of the unit except where
described in this manual as any unauthorized disassembly will
invalidate the warranty.
13. Before use on a patient, the oxygen concentration of the
delivered gas should be checked.
14. US Federal Law restricts this device to sale by or on the order
of a physician.
Minimum Flow rate
120 L/min
Exhalation Resistance
0 to 6 cmH
2
O @ 60 L/min
Inhalation Resistance
0 to -6 cmH
2
O @ 60 L/min
Oxygen Concentration
42.5 to 57.5%
Oxygen Enrichment Flow
(01EQ1000F only)
30 L/min
Input Pressure
50 to 70 PSI (3.5 - 4.8 Bar)
Low Input Pressure Alarm
40 ± 1 PSI (2.7 Bar)
Device Shut-off Pressure
35 ± 4 PSI (2.7 Bar)
Failsafe protection
~ 2 PSI diff erential pressure between O
2
and N
2
O
output gas. Prevents the output of high nitrous
oxide concentrations
Demand Valve Triggering
Pressure
-1.5 to -2.5 cmH
2
O
Supply Hose Connections
Refer to Section 6
Device Input Connections
Nitrous Oxide: CGA1040 DISS
Oxygen: CGA1240 DISS
Patient Connector
15/22 mm
Patient Circuit (Single use)
Dual limb circuit with one-way valves and
19 mm Scavenging Port Connection on the
exhalation limb.
Patient Valve Dead Space
32 ml
Operating Temperature
41
0
F to 104
0
F (5
0
C to 40
0
C)
Storage Temperature
-40
0
F to 140
0
F (-40
0
C to 60
0
C)
Weight
3.3 lb. (1.5 Kg)
Dimensions W x D x H
8.9”x6.6”x3.9” 226x168x99mm)