O-Two Equinox Relieve Скачать руководство пользователя страница 10

10

Equinox® Relieve - User Manual

Reverse gas fl ow test

To test the reverse fl ow, connect one gas at a time to the 

corresponding input connector, spray the other input connector, 

no bubbles allowed from this connector.

Function test

The following features of the Equinox® Relieve System can be 

individually tested or measured using a calibrated pressure gauge 

and fl owmeter during the Pre-use Functional Check:

1.  Demand Valve Function and Oxygen Concentration

Demand Valve function and oxygen concentration of the delivered 

gas can be determined by the following procedure: 

Note: 

To fully test this function it is necessary to have a supply 

regulator with an adjustable output pressure and a release valve 

(Not supplied). Checking of the alarm may be undertaken by 

simply slowly closing the cylinder valve.

A. Connect an oxygen monitor to the output connector of the 

Relieve using the “3-Way T” Connector test assembly supplied.

B. Utilizing a 1 Liter calibrated syringe, connect the syringe to 

the patient connection on the patient circuit connected to T- 

connector. Turn on the Relieve unit and activate the demand valve 

by cycling the syringe until the oxygen reading on the monitor 

has stabilized. Demand valve function is confi rmed by a change 

in the reading on the monitor.

Setting

Acceptable Range

50%

40% – 60%

2.  Low Input Pressure Alarms

With the adjustable inlet pressure regulator set to a supply 

pressure of 50 PSI, gradually reduce the supply pressure of the 

regulator to around 40 PSI while slowly releasing the gas until 

you hear the Low Input Pressure Alarm activated. For the nitrous 

oxide alarm the pulsed tone frequency is set at a low frequency 

of approximately 60 BPM. For the low oxygen alarm, the pulsed 

tone frequency is set at high frequency of approximately 120 BPM.

When low pressure alarm is activated, the corresponding Gas 

Supply Status Indicator will also cycle at the same rate of the alarm.

Содержание Equinox Relieve

Страница 1: ...Made in Canada by O Two Medical Technologies Inc Part Number 15PL2175 Rev 14 July 2023 USER MANUAL Equinox Relieve NITROUS OXIDE OXYGEN ANALGESIC GAS MIXING AND DELIVERY SYSTEM 01EQ1000 01EQ1000F...

Страница 2: ......

Страница 3: ...reparation for Use 2 1 Component List 2 2 Pre Use Functional Checks 3 Operating Instructions 3 1 Connection of Hoses 3 2 On Off Selector 3 3 Demand Valve 3 4 Low Input Pressure Alarms 3 5 Low Input Pr...

Страница 4: ...include but may not be limited to Hypersensitivity to the medication Head injuries with impaired consciousness Maxillofacial injuries Artificial traumatic or spontaneous pneumothorax Air embolism Midd...

Страница 5: ...I Both visual indicator and audible alarm circuits for both gasses are powered by oxygen only to prevent dumping nitrous oxide gas into atmosphere during alarm cycling The device is also equipped with...

Страница 6: ...od of two years from the date of purchase The manufacturer will replace and or repair all parts of the device at its option for two years from the date of purchase at no cost to the purchaser upon the...

Страница 7: ...s supply 6 It is recommended that an oxygen monitor be used with this device to monitor the gas output However if there is a concern that the device has not undergone regular maintenance at the prescr...

Страница 8: ...ut Pressure 50 to 70 PSI 3 5 4 8 Bar Low Input Pressure Alarm 40 1 PSI 2 7 Bar Device Shut off Pressure 35 4 PSI 2 7 Bar Failsafe protection 2 PSI differential pressure between O2 and N2 O output gas...

Страница 9: ...for CONNECTION OF HOSES ensure that the O Two Equinox Relieve ON OFF selector is in the OFF position and turn on the N2 O and O2 supplies Using a mild soap solution spray the input connections to the...

Страница 10: ...the Relieve using the 3 Way T Connector test assembly supplied B Utilizing a 1 Liter calibrated syringe connect the syringe to the patient connection on the patient circuit connected to T connector T...

Страница 11: ...to operate on medical nitrous oxide and medical oxygen from either individual gas cylinders or piped in medical gas systems The inlet fittings on the device are non interchangeable DISS fittings spec...

Страница 12: ...r inspiratory effort 3 4 Low Input Pressure Alarms Gas Supply Status Indicators Located on the front panel the O2 and N2 O visual indicators turn green when oxygen and Nitrous Oxide are supplied to th...

Страница 13: ...lieve System is equipped with a primary Failsafe protection circuit Should the output pressure differential of the internal mixer valve exceed 2 psi the Fail Safe protection circuit will activate a co...

Страница 14: ...activate instead of the low pressure alarm When the device is turned off the Failsafe protection circuit is deactivated Once initiated the alarm will continue its constant tone until the device is tur...

Страница 15: ...vice Service manual Meet product specifications Every 2 years Manufacturer service center Servicing Full service Service manual Meet product specifications Every 6 years Manufacturer Monthly inspectio...

Страница 16: ...as follows 1 Ensure that the device is disconnected from the gas supply source 2 Remove N2 O and O2 input hoses and wipe clean with a mild soap or hard surface disinfectant Ensure no cleaning solution...

Страница 17: ...oot 1 85 Meter O2 Supply Hose with AGA probe and 9 16 DISS Nut Device Connection 01FV4303 CZCH O Two 6 Foot 1 85 Meter O2 Supply Hose with CZECH probe and 9 16 DISS Nut Device Connection 01FV4303 DIN...

Страница 18: ...g and DISS Nut Ventilator Connection 01FV4303 DIN N2O O Two 6 Foot 1 85 Meter N2 O Supply Hose with DIN Nut and DISS Nut Device Connection 01FV4303 DISS N2O O Two 6 Foot 1 85 Meter N2 O Supply Hose wi...

Страница 19: ......

Страница 20: ...r In North America call Toll Free 1 800 387 3405 SERIAL N0 www otwo com 45A Armthorpe Road Brampton ON Canada L6T 5M4 Telephone 1 905 792 OTWO 6896 N A Toll Free 1 800 387 3405 Facsimile 1 905 799 133...

Отзывы: