14579-040 (PXIe System)
12
18 Slot PXIe Backplane (23007-518)
R1.5, January 2021
3.6 PXI Trigger Bus
The PXIe chassis trigger bus is divided into 3 isolated trigger bus segments with 8 trigger lines
each.
• Trigger segment 1 covers slot 1 through 6
• Trigger segment 2 covers slot 7 through 12
• Trigger segment 3 covers slot 13 through 18
Through the PXI trigger bridge, each trigger line of a segment can be individually routed in
either direction to the other segment.
The trigger bridges are controlled by the CMM.
1
2
3
4
6
7
8
5
H
H
10
11
12
9
H
H
Trigger
Segment I
Bus
P
XI T
ri
g
ge
r
B
ri
dg
e
Trigger
Segment I
Bus
I
CMM
Chassis Manager
I²C
H
H
H
16
17
18
H
H
P
XI T
ri
g
ge
r
B
ri
dg
e
Trigger
Segment I
Bus
II
H
13
14
15
H
H
H
I²C
H
H
H
H
By default the 3 trigger bus segments are isolated, the trigger bridges are disabled. To
enable the trigger bridges and set the direction of the routing, you have to access the
CMM‘s I²C register with the Schroff Embedded Controller or another appropriately
configured and programmed system controller.
For further information see CMM User Guide 63972-391
Содержание Schroff 14579-040
Страница 1: ...18 slot PXIe System User s Manual Product Number 14579 040 Doc No 63972 381_R1 5 January 2021...
Страница 4: ...14579 040 PXIe System II R1 5 January 2021...
Страница 29: ...14579 040 PXIe System R1 5 January 2021 Dimensions 25 8 Dimensions Figure 5 Dimensions...
Страница 30: ...14579 040 PXIe System 26 Dimensions R1 5 January 2021...
Страница 31: ......
Страница 32: ...Schroff GmbH Langenalber Str 96 100 75334 Straubenhardt Germany Tel 49 7082 794 0 Fax 49 7082 794 200...