50
14 Electromagnetic Compatibility (EMC) Information
1568913, Edition 2019-06, Version 0
en
14.1.2 Electromagnetic immunity
The Navigator Lift
™
180 is intended for use in the ELECTROMAGNETIC environment
specified below. The customer or the user of the Navigator Lift
™
180 should ensure that it
is used in such an environment.
Please note:
U
T
is the a.c. mains voltage prior to application of the test level.
Interference immunity test
Test level in accordance with IEC 60601
Test result
Electrostatic discharge in accordance with
IEC 61000-4-2
±8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
passed
Electrical fast transients / bursts in accor
-
dance with
IEC 61000-4-4
±2kV
100kHz repetition rate
passed
Surges
Line-to-line in accordance with
IEC 61000-4-5
±0.5kV, ±1kV
passed
Surges
Line-to-earth in accordance with
IEC 61000-4-5
±0.5kV, ±1kV, ±2kV
passed
Voltage dips in accordance with
IEC 61000-4-11
0% U
T
; 1/2 period
at 0, 45, 90, 135, 180, 225, 270 and 315 degrees
passed
0% U
T
; 1 period
and
70% U
T
; 25/30 periods
Single phase: at 0 degrees
passed
Voltage interruptions in accordance with
IEC 61000-4-11
0% U
T
; 250/300 periods
passed
Power frequency magnetic field immunity in
accordance with
IEC 61000-4-8
30A/m
50Hz or 60Hz
passed
Содержание Navigator Lift 180 Air Plus-Single
Страница 7: ...7 1568913 Edition 2019 06 Version 0 en...
Страница 56: ...No 1568913...