CLEANING SANITISATION DISINFECTION
Switch off the device before any cleaning procedure and unplug the power cable from the socket.
DEVICE AND TUBING EXTERIOR
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any
sort).
ACCESSORIES
Open the nebuliser by turning the upper part (C1.3) anticlockwise, remove the nozzle
(C1.2) from the upper part (C1.3) by pressing as shown by the 2 arrows in the "Connection
diagram" in point C1. Then proceed according to the following instructions.
SANITISATION
Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of
the methods described below.
method A: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 under potable hot
water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 in the dishwasher with
a hot cycle.
method C: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 by immersing them in
a solution of 50% water and 50% white vinegar, then rinse thoroughly under potable hot
water (approximately 40 °C).
If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
method A: Accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 by boiling them
in water for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 with a hot steam
steriliser for baby-bottle (not the microwave type). Perform the process faithfully following
the instructions of the steriliser. To ensure that the disinfection is effective, choose a
steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away
from dust.
REPLACING THE FILTER
The device has a suction filter (A3) that should be replaced when it is dirty or
changes color. Do not wash or reuse the same filter. Regular replacement of the
filter is required to help and ensure correct performance of the compressor.
To replace the filter:
Insert a flathead screwdriver between the edge of the filter and the body. Lift the
filter and remove it by turning and pulling it. The filter has been designed so that
it is always fixed in its seat. Do not replace the filter during use.
Only use original accessories and spare parts, we disclaim any liability in
the event of using non original spare parts or accessories.
TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
CAUSE
SOLUTION
The device does not work
The power cable has not been
correctly inserted into the socket
of the device or the power socket
Correctly insert the power cable
in sockets.
The device does not
nebulize or
nebulizes insufficiently.
The medication has not been
inserted in the nebulizer
Pour the right amount of
medication
in the nebulizer
The nebulizer has not been
properly fitted.
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on
the cover.
The nebulizer nozzle is
clogged
Disassemble the nebulizer,
remove the nozzle and perform
cleaning operations Failure to
remove medication deposits
from the nebulizer affects its
efficiency and operation.
Strictly comply with the
instructions contained in the
“CLEANING, SANITISATION,
DISINFECTION” chapter
The air pipe is not correctly
connected to the device
Ensure the device’s accessories
are properly connected to the air
intake (see connection diagram
on the cover).
The air pipe is bent, damaged
or twisted
Unwind the pipe and ensure it is
not crushed or punctured. If
necessary replace it.
The air filter is dirty
Replace the filter
The apparatus is noisier
than usual
The Filter is not correctly
inserted in its housing
Fully and properly insert the
Filter in its housing
If after verifying the above mentioned conditions the device should not operate
properly, contact your trusted retailer or the nearest authorized service center.