CLEANING SANITISATION DISINFECTION                                                                    

Switch off the device before any cleaning procedure and unplug the power cable from the socket.

DEVICE AND TUBING EXTERIOR

Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any 

sort).
ACCESSORIES

Open the nebuliser by turning the upper part (C1.3) anticlockwise, remove the nozzle 

(C1.2) from the upper part (C1.3) by pressing as shown by the 2 arrows in the "Connection 

diagram" in point C1. Then proceed according to the following instructions.
SANITISATION

Before and after each use, sanitise the nebuliser cup and the accessories, choosing one of 

the methods described below. 

method A: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5  under potable hot 

water (approximately 40°C) with a gentle, non abrasive dish detergent.

method B: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 in the dishwasher with 

a hot cycle.

method C: Sanitise accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 by immersing them in 

a solution of 50% water and 50% white vinegar, then rinse thoroughly under potable hot 

water (approximately 40 °C).

If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION 

paragraph.

After having sanitised the accessories, shake them vigorously and place them on a paper 

towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer). 

DISINFECTION

After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the 

methods described below. 

method A: Accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 can be disinfected.

The disinfectant must be an electrolytic chloroxidizer (active principle: sodium 

hypochlorite) specific for disinfecting, which is available in any pharmacy.

Implementation:

-   Fill a container big enough to hold all of the parts to disinfect with a solution of potable 

water and disinfectant, according to the proportions indicated on the packaging of the 

disinfectant.

-   Completely immerse each part in the solution, taking care to avoid the formation of air 

bubbles on the parts. Leave the parts immersed for the amount of time indicated on the 

packaging of the disinfectant associated with the concentration chosen for the solution.

-   Remove the disinfected parts and rinse abundantly with warm potable water.

-  Dispose of the solution following the instructions provided by the disinfectant 

manufacturer.

method B: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 by boiling them 

in water for 10 minutes; use demineralised or distilled water to prevent calcium deposits.

method C: Sanitise the accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 with a hot steam 

steriliser for baby-bottle (not the microwave type). Perform the process faithfully following 

the instructions of the steriliser. To ensure that the disinfection is effective, choose a 

steriliser with an operating cycle of at least 6 minutes.

After having disinfected the accessories, shake them vigorously and place them on a 

paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).

At the end of each use store the device complete with accessories in a dry place away 

from dust.

REPLACING THE FILTER                                                                                                       

The device has a suction filter (A3) that should be replaced when it is dirty or 

changes color. Do not wash or reuse the same filter. Regular replacement of the 

filter is required to help and ensure correct performance of the compressor.  

To replace the filter: 

Insert a flathead screwdriver between the edge of the filter and the body. Lift the 

filter and remove it by turning and pulling it. The filter has been designed so that 

it is always fixed in its seat. Do not replace the filter during use.

Only use original accessories and spare parts, we disclaim any liability in 

the event of using non original spare parts or accessories.

TROUBLE-SHOOTING                                                                                                        

Switch off the device before any procedure and unplug the power cable from the socket.

PROBLEM

CAUSE

SOLUTION

The device does not work

The power cable has not been 

correctly inserted into the socket 

of the device or the power socket

Correctly insert the power cable 

in sockets.

The device does not 

nebulize or

nebulizes insufficiently.

The medication has not been 

inserted in the nebulizer

Pour the right amount of 

medication

in the nebulizer

The nebulizer has not been 

properly fitted.

Disassemble the nebulizer and

reassemble it correctly as shown

in the connection diagram on 

the cover.

The nebulizer nozzle is 

clogged

Disassemble the nebulizer, 

remove the nozzle and perform 

cleaning operations Failure to 

remove medication deposits 

from the nebulizer affects its 

efficiency and operation.

Strictly comply with the 

instructions contained in the 

“CLEANING, SANITISATION, 

DISINFECTION” chapter

The air pipe is not correctly 

connected to the device

Ensure the device’s accessories

are properly connected to the air

intake (see connection diagram 

on the cover).

The air pipe is bent, damaged 

or twisted

Unwind the pipe and ensure it is

not crushed or punctured. If 

necessary replace it.

The air filter is dirty

Replace the filter

The apparatus is noisier 

than usual

The Filter is not correctly

inserted in its housing

Fully and properly insert the 

Filter in its housing

If after verifying the above mentioned conditions the device should not operate 

properly, contact your trusted retailer or the nearest authorized service center.