33
Error 02:
Air Detector –
During sterilization, difference between two temperature sensors in
the chamber is more than 3 degree.
Error 03:
Steam Discharge –
The steam in the chamber is not released within the preset time
following the sterilization phase.
Error 04:
Air intake –
Following the drying phase, the airflow rate is below the preset value.
Error 06:
Door Open –
Door lock has been released during operation.
Error 07:
Steam Generator Over Heating –
Steam generator is overheated.
Error 10:
Sensor Failure PT1, PT2, PT4, BT2 –
Sensor is broken
.
Error 11:
Pre-Heating –
The duration for pre-heating has exceeded the preset time.
Error 12:
Pre-Heating High Temperature –
The maximum pre-heating temperature has been
exceeded.
Error 13:
Water Pump Failure –
Water pump is defective or filling pipe is clogged.
Error 14:
Door Lock Failure –
Door lock activating time exceeds the time allowed after the
program beginning.
Error 16:
Power Failure –
The mains supply has interrupted during a cycle. If chamber is
pressurized in case of power failure, pressurized air is released. Thus, the filter on the air release
line should be sterilized by running quick program after power failure.
In case of power failure during sterilization phase, chamber temperature is checked
when the power is on again. Program continues to run if the temperature is in the
safe limits. Otherwise, it stops. If the power failure occurs before or after the
sterilization phase, the operation stops and audible and visual alarm appears.
Error 19:
Low Temperature –
The temperature in the chamber remains below the preset
temperature after the sterilization phase has started.
Error 20:
High Temperature –
The sterilization temperature exceeds the maximum permitted
temperature.
Error 21:
Low Pressure –
The pressure in the chamber remains below the preset pressure after
the sterilization phase has started.
Error 22:
High Pressure –
The sterilization pressure exceeds the maximum permitted pressure.
Error 23
:Door Lock Cannot Open -
This message appears on the screen when door lock could
not open. Ensure that lid handle is closed properly.
Error 30:
Steam supply –
This message appears when required for temperature and pressure
values are not reached at Preheating phase which is preparation phase before sterilization.