TEST PARAMETERS
CONTROL RANGE
1.
Check the safety valve by operating it.
2 months
2.
Remove the cover of the autoclave, check and tighten the
ports and valves.
6 months
3.
Check the lid gasket.
6 months
4.
Check that the autoclave is leveled.
Annual
5.
Check the continuity of the grounding connections.
Annual
6.
Check the safety elements (safety valve, safety and cut-off
thermostats and lid locking mechanism).
Annual
7.
Check the water reservoir, piping and plastic parts of
autoclave.
Annual
8.
Run the sterilization programs of autoclave and check the
operational/ sterilization parameters.
Annual
9.
Check the precise operation of the earth leakage relay and
electrical control systems.
Annual
10.
Check and tighten all screw connections in the control box,
valves and instrument.
Annual
11.
Check the temperature sensor calibration.
Annual
12.
Validate autoclave effectiveness (loading/ unloading).
6 months/ Annual
13.
Observe the closing device for excessive wear.
5 years
14.
All safety valves exposed to direct steam pressure must be
checked.
5 years
Safety tests (pressure vessel, efficiency, electrical) shall be performed in accordance
with local rules or regulations, by an authorized inspector.
According to calculations, number of allowable cycles for the operation conditions are
10.000 at pressure fluctuation between 0 bar to 2,05 bar and 20.000 at pressure
fluctuation between 0 bar to 1,05 bar.
8.3.
CLEANING
Weekly cleaning should be performed if the sterilizer operates daily. Use liquid detergent to
remove tough dirt. Take precautions while handling chemical cleaners. Please be aware of
the undesirable effects of the chemicals and be careful while applying them.
A soft washcloth shall be used not to cause any detriment in the chamber.
The chamber shall be checked before loading sterilizer; and shall be immediately cleaned if
needed.
The sterilization load should be disinfected prior to loading sterilization chamber.