3.4 MAINS SUPPLY
The incubator requires 230 V, 50 Hz.
Please make sure that the supplied mains matches the required power ratings.
Always plug-
in the ıncubator to correctly grounded sockets.
3.5 PRIOR TO INCUBATION
Plug the instrument in to a grounded socket.
Check the followings,
Make sure that the safety thermostat is adjusted to the temperatures which are higher than
the set temperature.
If it is necessary the ventilation hole is open to discharge the gases and the vapours which
occur during incubation.
Liquids are not heated in sealed containers.
The boiling points of the samples are higher than the set temperature.
Liquids which may expand during heating do not overflow from their containers.
The vapours and gases which are generated during the operation are not harmful to humans
or flammable or explosive.
The set temperature does not destroy the structure of the samples.
Plug the power cable into a grounded socket.
The safety thermostat set value should always be set to a value which higher than the
working temperature.
NOTE :
Never use explosive, flammable, acidic or toxic liquids. Read carefully the functions of
the control panel.
SECTION 4
OPERATING PRINCIPLE
4.1 SWITCHING ON
Push On/Off switch.
See that the microprocessor control system activates.
Set the values and start the operation.
A supply fitted with a circuit breaker should be used for protection against indirect
contact in case of an isolation fault.
Содержание EN 300
Страница 14: ...13 SECTION 8 ELECTRICAL CIRCUIT DIAGRAM 8 1 EN 300 ELECTRICAL CIRCUIT DIAGRAM...
Страница 15: ...14 8 2 EN 400 ELECTRICAL CIRCUIT DIAGRAM...
Страница 16: ...15 8 3 EN 500 ELECTRICAL CIRCUIT DIAGRAM...
Страница 17: ...16 8 4 EN 400P ELECTRICAL CIRCUIT DIAGRAM...
Страница 18: ...17 8 5 EN 500P ELECTRICAL CIRCUIT DIAGRAM...
Страница 19: ...18 SECTION 9 WARNING LABEL...
Страница 20: ...19 GROUNDED PLUG EN 300 FUSES 2x1A EN 400 EN 400P FUSES 2x 1 6A EN 500 EN 500P FUSES 2x2A...