Deflating

Deflating

To a fit pressure

Inflating

Start to inflate

13

In order to accelerate the measurement time for the systolic pressure more 

than 180 mmHg, press and hold the memory button until the unit inflates 30 to 

40 mmHg higher than your suspected systolic pressure if you know. ( The unit 

will not inflate above 299 mmHg).

INSTRUCTIONS FOR SPECIAL CONDITIONS

4

Press the “START/STOP” button to turn off the unit.

Take a Measurement

2

The heartbeat symbol flashes once a pulse is detected.

The unit will anutomatically turn off after 2 minutes if you have 

forgotten to turn it off.

The cuff is deflated automatically, 

and all of the measurement 

results are stored in the memory .

The “    ” symbol will be displayed 

if 

irregular heartbeat

 

is detected.

3

The blood pressure and pulse rate are displayed when 

the measurement completed.

All display symbols appear on the screen. The cuff starts to 

inflate automatically.

Press the “START/STOP” button.

1

Measurement starts while inflating.

NOTE: To stop the inflation or measurement,push the “START/STOP” 

              button,The unit will stop inflating and then release and turn off.

All symbols display

Start to inflate

Deflating automatically

OR

"mmHg" display

"kPa" display

OR

"     " display

12

Proper Use of the Arm Cuff

8

Relax your arm, keep the 

palm upward and the 

fingers curving naturally, 

then turn on the unit and 

start to measure.

Wrap the cuff tightly around 

your upper arm by using the 

Velcro strip. Do not leave 

any spacing between the 

arm and the cuff.

7

1 ~ 2cm

The hose runs down the 

inside of your arm. The 

bottom of the cuff should be 

approximately 1~2cm above 

your elbow.

6

5

Put your arm through the 

loop, then pull it up to the 

position of your upper arm.

BA2110

Содержание BA2110

Страница 1: ...Model No BA2110 VER A00...

Страница 2: ...t 4 5 IHB Introduction 6 Before Taking A Measurement 7 Quick Reference Guide 8 Battery lnstallation Replacement 9 Adjusting Settings 10 Proper Use of the Arm Cuff 11 12 Take a Measurement 13 Use the M...

Страница 3: ...or moderate injury to the user or patient or damage to the equipment or other property Read all information in the instruction manual and any other literature included in the box before using the unit...

Страница 4: ...ff A new arm cuff does not include an air plug Please continue to use the old air plug on the new arm cuff Hose Air Plug Insert to Air Socket Metal Arm Cuff Warning 1 Do not use rechargeable batteries...

Страница 5: ...d bathing for 30 minutes before taking a measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 The cuff can be applied to your left or right wrist Always measur...

Страница 6: ...tive polarities match the polarities of the battery compartment as indicated Remove the battery cover from the battery compartment Replace the battery cover Press the START STOP button All symbols app...

Страница 7: ...heart 3 Remove all clothes from your upper arm allowing the cuff to fit directly on the skin 2 Make sure the air plug is inserted properly in the main unit 1 10 Press the MEM button about 10 seconds u...

Страница 8: ...3 The blood pressure and pulse rate are displayed when the measurement completed All display symbols appear on the screen The cuff starts to inflate automatically Press the START STOP button 1 Measur...

Страница 9: ...d to save the latest measurement value when more than 60 sets The unit also calculates an average reading based on the values of the latest 3 times measurement taken To delete the memory value In the...

Страница 10: ...nts the blood pressure produced by relaxation of the heart muscle Blood Circulation 17 About Blood Pressure Note When a person s systolic and diastolic pressures fall into different categories the hig...

Страница 11: ...ve your arm and body and keep quiet Error Indicators 18 The following symbol will appear on the display when measuring abnormally 19 Care and Maintenance Do not clean the body and cuff with naphtha th...

Страница 12: ...c and heart rate from the upper arm by using the oscillometric method The unit is intended for usingin only adult population not applied to the other populations such as neonatal baby It can t be used...

Страница 13: ...60601 1 2 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic compatibility Requirements and tests EN1060 1 Non in...

Страница 14: ...1000 4 8 6 kV contact 8 kV air 6 kV contact 8 kV air 3A m 3A m Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at least 30 P...

Страница 15: ...be necessary such as reorienting or relocating the Digital Blood Pressure Monitor Radiated RF IEC 61000 4 3 IEC 60601 test Compliance level level Electromagnetic environment guidance Recommended sepa...

Отзывы: