OM0246
Page 7 of 64
Rev 5 June/2020
3.0
Models & Features
3.1
Invitrocell DHD (heated decontamination) all Models.
3.2
Dimensions (see also Specification Drawing BCD-18314)
Overall Dimensions - inches (mm):
Model NU-5810/E thru NU-5841/E
Height:
Exterior: 39.688 in (1008 mm)
Width:
26.750 in (680 mm)
Depth:
27.188 in (691 mm)
Foot print:
23.5 in (597 mm) wide x 20.75 in (527 mm) deep
Height:
Interior: 28.500 in (723.900mm)
Width:
21.257 in (539.928mm)
Depth:
20.053 in (509.346mm)
Shelf Capacity:
Size:
19.75 Inches (502mm) x 18.75 Inches (476mm)
Supplied:
4 Shelves
Maximum Capacity:
20 Shelves
Maximum Weight Capacity
25 lbs. per DHD shelf
125 lbs. per DHD Incubator
(Make sure self is empty before changing position or removing.)
Water Pan:
Dimensions:
Mean Length
12.0 inches (305mm)
Mean Width
10.0 inches (254mm)
Depth
1.5 inches (38mm)
Capacity:
Maximum
2.0 Liters
Recommended Fill
1.5 Liters
3.3
Performance Parameters :
Temperature Control: Available on All Models
Performance Determined with O₂ Set Point @ 21.0%
•
Temperature Range:
5
°
C to 55
°
C (37°C default) (5
°
C above ambient to 30
°
C maximum ambient)
•
Temperature Uniformity:
+0.3°C @ 37°C.
•
Temperature Accuracy:
+0.1°C.
•
Temperature Recovery:
0.12°C/minute Average
•
Temperature Display Resolution:
0.1°C
•
Temperature Sensor Type:
Precision Integrated Circuit
•
Minimum Qualifications for
Decontamination/Sterilization:
145 DEG Cycle:
95 DEG Cycle
Temperature:
135°C
87°C
Time:
2 Hr.
9 Hr.
•
Door and Perimeter:
Proportional base duty cycle based on Temperature set point
and -20 to +20% manually adjustable to adapt to ambient conditions.
CO
2
Control: Available on All Models
Performance Determined with O₂ Set Point @ 21.0%
•
CO
2
Range:
0.1 to 20% (default 5%). (0.0 set point idles system)
•
CO
2
Accuracy:
+0.1%
•
CO
2
Recovery:
Up to 5% +0.2/-0.5 in 5 minutes average.
•
CO
2
Display Resolution:
0.1%
•
CO
2
Control Logic:
Fixed Algorithm/Manual Environmental Adaptable.
•
CO
2
Sensor Type:
Infrared single source dual wave length
•
Sensor Bay Temp
Chamber Set Point +3.5 Degree C.
Содержание NU-5810
Страница 9: ...OM0246 Page 9 of 64 Rev 5 June 2020 ...
Страница 13: ...OM0246 Page 13 of 64 Rev 5 June 2020 ...
Страница 26: ...OM0246 Page 26 of 64 Rev 5 June 2020 8 7 Run Mode Operator Interactions ...
Страница 58: ...OM0246 Page 58 of 64 Rev 5 June 2020 ...
Страница 59: ...OM0246 Page 59 of 64 Rev 5 June 2020 ...
Страница 60: ...OM0246 Page 60 of 64 Rev 5 June 2020 ...
Страница 61: ...OM0246 Page 61 of 64 Rev 5 June 2020 ...
Страница 62: ...OM0246 Page 62 of 64 Rev 5 June 2020 ...
Страница 63: ...OM0246 Page 63 of 64 Rev 5 June 2020 ...
Страница 64: ...OM0246 Page 64 of 64 Rev 5 June 2020 ...