background image

Document: SUII02 Rev.7 

 

 

 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

Description of Equipment: 
Beschreibung des Gerätes: 
Description du matériel: 
Descripción de la máquina:

 

OxyProbe200 - Oxygen Sensors For Use in Hazardous Areas. 
OxyProbe200 - Sauerstoff-Sensoren für den Einsatz in 
explosionsgefährdeten Bereichen/OxyProbe200 - capteurs d'oxygène pour 
une utilisation dans des zones dangereuses/ OxyProbe200 - Los sensores de 
oxígeno para su uso en áreas peligrosas. 

 

Standards applied: 
Normes appliquées: 
Normas aplicadas: 

 
 
 

EN60079-0:2012 + A11:2013 
EN60079-11:2012 
 

 

Marking: 
Kennzeichnung: 
Marquage: 
Marcado:

 

 

          II   

II 1 GD      Ex ia IIC T6 Ga(-20°C≤Ta≤+55°C) 

                           Ex ia IIIC T90°C Da( -20°C≤Ta≤+55°C) 
                           IECEx BAS 09.0148X 
Certificate No:   BAS02ATEX1230X  

 

 

Notified Body: 
Benannte Stelle: 
Organisme notifié: 
Organismo notificado: 

 

SGS Baseefa 
Rockhead Business Park 
Staden Lane, Buxton, Derbyshire, 
SK17 9RZ, UK. 
Number/Anzahl/nombre/número: 1180 
 
 

 
The Authorised Signatory to this declaration, on behalf of the manufacturer, is identified below: 

Der Prokurist dieser Erklärung im Namen des Herstellers, ist unten angegeben: 
Le signataire autorisé à cette déclaration, le nom du fabricant, est identifié ci-dessous: 
El signatario autorizado a esta declaración, en nombre del fabricante, se identifica a continuación: 
 

Name:

   David Beirne 

 

Title/Titel/ Titre/ título:

  Managing Director/Geschäftsführer/ Directeur-        

Général/ director Gerente 

Address/Adresse/ dirección:

 

Ntron Limited 

 

 

 

 

Mullaghboy Industrial Estate 

 

 

 

 

Navan 

 

 

 

 

Co. Meath 

 

 

 

 

Ireland, C15 XD61 

 

Signature: 

 

 

 

 

Date: 

25/02/2019 

EU Declaration of Conformity 

EG-Konformitätserklärung / Déclaration de conformité CE / 

Declaración de conformidad CE: 

 
We declare, under our sole responsibility, that we believe the products identified in this declaration, and to which this 
declaration relates are in conformity with the requirements of the EU Council Directives as stated below; 
Directive 2014/34/EU equipment and protective systems intended for use in potentially explosive atmospheres. 
Directive 2011/65/EU and of Council of 8th June 2011 on the restriction of the use of certain hazardous substances in 
electrical and electronic equipment (RoHS2)(*) 
*OJ L 174, 1.7.2011, p.88 & P106 Annex IV 1b exemption. 

Wir erklären in alleiniger Verantwortung, dass wir der Ansicht sind, dass die in dieser Erklärung genannten Produkte, auf die sich diese Erklärung 
bezieht, den Anforderungen der unten aufgeführten Richtlinien des EU-Rats entsprechen; Richtlinie 2014/34 / EU Geräte und Schutzsysteme zur 
bestimmungsgemäßen Verwendung in explosionsgefährdeten Bereichen.Richtlinie 2011/65 / EU und des Rates vom 8. Juni 2011 zur Beschränkung 
der Verwendung bestimmter gefährlicher Stoffe in Elektro- und Elektronikgeräten (RoHS2) (*)* ABl. L 174 vom 1.7.2011, S. 88 & S. 106 Anhang 
IV 1b Befreiung.  
Nous déclarons, sous notre seule responsabilité, que nous croyons que les produits identifiés dans cette déclaration, et auxquels cette déclaration se 
rapporte, sont conformes aux exigences des directives du Conseil de l'UE comme indiqué ci-dessous;Directive 2014/34 / EU équipements et 
systèmes de protection destinés à être utilisés dans des atmosphères explosibles.Directive 2011/65 / UE et du Conseil du 8 juin 2011 relative à la 
limitation de l'utilisation de certaines substances dangereuses dans les équipements électriques et électroniques (RoHS2) (*) * JO L 174 du 1.7.2011, 
p.88 et P106 Annexe IV 1b. 
Declaramos, bajo nuestra exclusiva responsabilidad, que creemos que los productos identificados en esta declaración, y a los que se refiere esta 
declaración, están en conformidad con los requisitos de las Directivas del Consejo de la UE como se establece a continuación; Directiva 2014/34 / 
UE equipos y sistemas de protección destinados a ser utilizados en atmósferas potencialmente explosivas. Directiva 2011/65 / UE y del Consejo, de 
8 de junio de 2011, sobre la restricción del uso de determinadas sustancias peligrosas en aparatos eléctricos y electrónicos (RoHS2) (*) * DO L 174 
de 1.7.2011, p.88 y P106 Anexo IV 1b exención.

 

 

Содержание SIL O2

Страница 1: ...1 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 The SIL O2 Oxygen Analyser User Operation Manual Rev 1 1 30 11 2018...

Страница 2: ...2 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8...

Страница 3: ...3 3 Typical Relay Configuration 13 2 3 4 Sensor Process Connection options 15 2 4 Operation 16 2 4 1 Calibration Procedures 16 2 4 2 Setting the Zero point 17 2 4 3 SPAN Calibration 18 18 3 Operationa...

Страница 4: ...vision History Revision No Change Description Date 0 First Issue 19 10 2018 1 Note added for Non ATEX Sensor usage 22 10 2018 1 1 Reference to 15 O2 in trouble shooting section removed 30 11 2018 Cali...

Страница 5: ...5 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8...

Страница 6: ...et according to the end user s requirements Interface software and hardware is available for the user to perform such functions themselves Contact Ntron for further details 1 1 The purpose of this Use...

Страница 7: ...embly disassembly installation operation and maintenance may only be performed by qualified personnel in accordance with regulatory requirements and the SIL O2 Oxygen Analysis Manual During installati...

Страница 8: ...8 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 The Ntron Sensors suitable for use with the SIL O2 analyser are listed in the Installationn section which follows...

Страница 9: ...ote Non Atex Ex Sensors variants of the models listed in the manual may also be used with the SIL O2 Analyser as long as they are installed in a Non ATEX Ex or Safe area or zone For Acetal bodies Sens...

Страница 10: ...Sensor types and related process connections are given below Sensor Type Process Connection Model OC 25 Ntron Sensor base Flow Through or Tri Clamp Model OC 26 KF40 Flange Model OC 200 Oxyprobe series...

Страница 11: ...rameter values which match the Isolation Barrier output which is built into the SIL O2 Analyser The Sensor connects to the Blue terminals as shown below The dark Grey terminals are for safe Non Ex are...

Страница 12: ...ose when power is applied and the Analyser and Sensor are healthy and ready to operate The Analogue output is an industry standard 4 20mA active source output The connected circuit should be of maximu...

Страница 13: ...customer user when the SIL O2 module is under power and the status is healthy good When at a level setpoint Alarm or when the SIL O2 module is not powered the relays are de energised not under current...

Страница 14: ...14 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8...

Страница 15: ...ble for mounting directly onto process Lines via a selection of process fittings The electrical signal connectors are rated to IP67 Tri Clamp Sensor Base Flow Through Sensor Base OC 25 Sensor OC 26 Se...

Страница 16: ...m or fault events and their conditions please see the troubleshooting section at the end of this Manual 2 4 1 Calibration Procedures The SIL O2 Analyser and Sensor unit requires periodic calibrations...

Страница 17: ...2 unit will generally give a small output at zero oxygen levels The zero adjustment function allows the user to adjust this small signal so that the SIL 2 analyzer displays 0 0 when no oxygen is prese...

Страница 18: ...9 oxygen If an ambient air calibration is performed it is recommended to confirm the oxygen level with a certified portable oxygen analyzer prior to calibration If a zero setting is made first then a...

Страница 19: ...w seconds following an automatic internal cyclic check 3 2 Wire break When manually disconnecting the sensor for example for maintenance or following a break in the sensor connection cable the SIL O2...

Страница 20: ...yser with no sensor connected or a wire break to the sensor Problem If the sensor is disconnected this would ordinarily cause relays R1 R2 and R3 to de energise Analyser display and Analogue output wi...

Страница 21: ...cuit is broken or interrupted Investigate and repair 3 4 1 Possible Sensor Faults Applicable to all sensor types OC 25 OC 26 OC 200 Problem SIL O2 Analyser will not calibrate Cannot be set to 20 9 Sol...

Страница 22: ...or used has a finite life span The Performance of the Sensor is verified by calibration check and this should be performed according to the protocols required by the safety system Replacement of the S...

Страница 23: ...IEC 947 5 1 resp EN 60947 Transistor Output Do Switching parameters 28V 50mA Intrinsically Safe Connection voltage Uo DC 6 V current intensityIo 0 2 mA power Po 0 3 mW max outer inductivity Lo 1000 m...

Страница 24: ...C 26 OC 200 series Range 0 25 Oxygen Signal Output 300 375mV in Air OC25 OC 26 135 160mV in Air OC 200 Technology Lifespan Electrochemical Solid State Long Life 3 5 Years application dependant Storage...

Страница 25: ...ase OC 26 KF40 Flange OC 200 Probe Holder mechanism or Bushing Weight OC 25 250g OC 26 150g OC 200 180g Environmental Temperature 20 to 45 50 C 10 95 Humidity no condensation Certification Standards C...

Страница 26: ...26 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 5 Appendices SIL O2 Analyser and Sensor Connection Diagrams SIL O2 Analyser CE and ATEX Certificates Sensor CE and ATEX Certificates...

Страница 27: ...27 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 5 1 SIL O2 Analyser and Sensor Connection Diagrams E364 SIL O2 with OC 25 Sensor E511 SIL O2 with OC 26 Sensor E395 SIL O2 with OC 200 Oxyprobe...

Страница 28: ...Gas Measurement...

Страница 29: ...Gas Measurement...

Страница 30: ...Gas Measurement...

Страница 31: ...Gas Measurement...

Страница 32: ...Gas Measurement...

Страница 33: ...28 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 5 2 SIL O2 Analyser CE and ATEX Certificates CE Certificate CETX003 ATEX Certificate Dekra BVS 13 ATEX E 010...

Страница 34: ...hen Directive 2014 34 EU quipements et syst mes de protection destin s tre utilis s dans des atmosph res explosibles Directiva 2014 34 UE equipos y sistemas de protecci n destinados a ser utilizados e...

Страница 35: ...Organismo notificado Organismo notificato DEKRA Exam GmbH Dinnendahlstrasse 9 44809 Bochum Germany BVS 13 ATEX E 010 The Authorised Signatory to this declaration on behalf of the manufacturer is ident...

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ...29 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 5 3 Sensor CE and ATEX Certificates CE Certificate SUII01 SUII02 ATEX Certificate BAS02ATEX1230X 11...

Страница 47: ...ipment RoHS2 OJ L 174 1 7 2011 p 88 P106 Annex IV 1b exemption Wir erkl ren in alleiniger Verantwortung dass wir der Ansicht sind dass die in dieser Erkl rung genannten Produkte auf die sich diese Erk...

Страница 48: ...une journ e Si le capteur ne se r talonne plus il doit tre remplac GERMAN BEDIENUNGS UND INSTALLATIONSANLEITUNG DES SENSORS Um die Sicherheit von Personal G tern und Anlage gew hrleisten zu k nnen ist...

Страница 49: ......

Страница 50: ...recalibrer il doit tre remplac Il est recommand que les capteurs tre remplac s apr s 36 mois GERMAN BEDIENUNGS UND INSTALLATIONSANLEITUNG DES SENSORS VORABHINWEIS Um die Sicherheit von Personal G ter...

Страница 51: ...ems intended for use in potentially explosive atmospheres Directive 2011 65 EU and of Council of 8th June 2011 on the restriction of the use of certain hazardous substances in electrical and electroni...

Страница 52: ......

Страница 53: ......

Страница 54: ......

Страница 55: ......

Страница 56: ...30 P a g e R e v 1 1 I s s u e 3 0 1 1 2 0 1 8 Ntron Ltd Mullaghboy Industrial Estate Navan County Meath Ireland 00353 46 9071333 www ntron com...

Отзывы: