English
19
Water does not exit, or flow
level is low.
Foot Control is not connected.
Check the connection.
Irrigation Tube is not connected to Irrigation
pack or Handpiece.
Check the connection.
Irrigation Tube is not properly mounted to
the pump.
Check the connection.
The Pump Cover is open.
Close the cover.
The Irrigation Tube is broken (leaking).
Replace the Irrigation Tube.
Irrigation leak.
Leak from the connection between
Irrigation pack and Irrigation Tube.
Insert Irrigation Tube Needle all the way
into the Irrigation pack.
Leak from the connection between
Handpiece and Irrigation Tube.
Insert the end of the Irrigation Tube
completely into the Handpiece water pipe.
Leak from Irrigation Tube.
Replace the Irrigation Tube.
Irrigation does not stop.
Mode is set to Auto Cleaning.
To stop irrigation, press the Flow key.
Faulty circuit.
Service only by Authorized NSK Dealer.*
Handpiece LED does not
illuminate.
LED is burned out.
Service only by Authorized NSK Dealer.*
Failure inside the Control Unit or inside the
Handpiece Cord.
Service only by Authorized NSK Dealer.*
Failure or disconnection of the circuit.
Service only by Authorized NSK Dealer.*
LCD display does not properly
display.
Part of display is missing.
Overheating.
Heat can cause LCD display malfunction.
Turn the Main Power OFF to cool down.
Error Code is displayed.
Refer to the Error Code table.
Failure of the LCD display or driving circuit. Turn the Main Power OFF and then ON. If
part of initial display is still missing
Contact your Authorized NSK Dealer.
Settings are not memorized
Wrong settings are
memorized.
Figures have been temporarily changed.
Increase or decrease the Program No. by
one, and then return to the original setting.
Foot control is being operated.
Settings cannot be saved while LED
Handpiece with cord is operating.
An Error Code is displayed.
Refer to the Error Code table.
* Repairs cannot be made by the customer.