NSK Varios 560 Скачать руководство пользователя страница 7

10

11

Autoclave Procedure

1) Remove the Tip after use. (Refer. 4. (2)  Tip Connection)
2) Remove the handpiece from the Handpiece Cord. (Refer. 3. (5)  Handpiece Connection)
3) Wipe off dirt and debris on the handpiece with an alcohol soaked cloth.
4) Insert into an autoclave pouch. Seal the pouch.
5) Autoclavable up to max. 135°C.

ex.) Autoclave for 20 min. at 121°C, or 15 min. at 132°C.

6) Keep the products in the autoclave pouch to keep it clean until you us.

    Sterilization at 121°C for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.

WARNING ON STERILIZATION

· This products can not be cleaned and disinfected with a Thermo-Disinfector.
· Do not sterilize by ultraviolet ray. The handpiece could discolor.
· If autoclaved with other instruments stained with chemical solution, it could strip the plating and make the 

surface black.

· Since the bottom shelf of the sterilizer is close to the heat source, it can be too hot. Place instruments in the 

middle or top shelves.

· Do not autoclave any parts (the Control Unit, Power Cord, Foot Control, Water Tube, Water Connector, Handpiece 

Cord including the Cover, Lamp, O-ring, and Spanner Wrench (5 X 8) ). Other than those that can be subjected to 
autoclave sterilization. Perform alcohol disinfection to the Control Unit, Power Cord, Foot Control, Handpiece Cord 
including with reference after every patient.

· Do not sterilize diamond coated Scaler Tips for the reason of single use.

(1) How to use the Tip Card

1) Place the neck of the Tip in the cut out.
2) Check wear of the Tip.
3) See the green, yellow and red line to check wear of the Tip. *See below what each color means. At NSK we 

recommend to replace a Tip when the Tip meets the yellow line (wear of 1mm) to guarantee safe and effective use.

8. Care and Maintenance

Green: No wear - Tip is OK

Tip replacement is not necessary.

Yellow: Wear of 1mm - Tip is showing some wear

Tip replacement is recommended.

Red: Wear of 2mm - Tip is badly worn

Tip replacement is necessary.

Fig. 20

OK

NO

Tip Card

Handpiece

Neck

*The Tip Card can be used to check the following tips : G1, G4, 

G5, G6, G8, P1/P1D, P10, and P20

Efficiency

1mm

2mm

25%

50%

Decrease

Decrease

(2) Changing Water Filter

1) Close the water valve of the dental unit to which Varios 

560 LUX or Varios 560 is connected. Mount two 

Spanner Wrenches(5 X 8) as shown in Fig. 21, and 

turn in the directions shown.

Fig. 21

2) When the Water Filter Case is separated, the Water Filter 

can be removed as shown in Fig. 22. Replace with a new 
one and reassemble the Filter in the reverse order.

Optional Water Filter : Order No.U387042

Fig. 22

When replacing the water filter, be careful about its 
direction. (Point the O-ring on the water filter toward 
the control unit side.) If it is assembled in the opposite 
direction, the effect of the water filter will be lost.

CAUTION

CAUTION

NOTICE

Tips are consumables. The efficiency of dental scaling 
decreases approximately 25% when the top of the Tip 
wears 1 mm and approximately 50% when it wears 2 
mm. In addition, the vibration condition changes owing 
to the wear, which may damage a patient’s tooth 
surface. Check the Tip wear condition with the Tip Card 
periodically, and replace the Tip with a new one in good 
time.

When Water Tube is getting twisted by this, relieve the twist by turning it, as its end on the Control Unit side is free 
to turn.

CAUTION

(3) Cleaning of Donut-Shape Light (Varios 560 LUX)

Wipe the debris off the End Face Optic Fibers at the 

handpiece with alcohol soaked cotton swab.

Fig. 23

Do not use any sharp pointed tools to clean the End 
Face Optic Fiber. In case the light becomes dim, 
contact your dealer.

Water Filter Case

Water Filter Case

Water Filter

Water Filter

O-ring

O-ring

Control Unit Side

Control Unit Side

Cotton Swab

Cotton Swab

End Face Optic Fiber

End Face Optic Fiber

Содержание Varios 560

Страница 1: ...read this Operation Manual carefully before use and file for future reference Multi Function Ultrasonic Scaler OM E0248E 004 10 03 01 S The EU directive 93 42 EEC was applied in the design and production of this medical device ...

Страница 2: ...evice Safety instructions are classified as follows in accordance with the seriousness of the risk Use by medical professional such as doctor or dental hygienist is intended Do not unplug the Power Cord with wet hands to avoid electric shock Be sure to prevent water on the Control Unit because it may result in short circuit and electric shock Do not allow any impacton the Control Unit Do not drop T...

Страница 3: ...on and overheating outside the patient s oral cavity before use If any abnormalities are found stop using immediately and contact your dealer Should this products function abnormally cease operation immediately and returen it to the dealer for repair Do not force or pull on Power Cord and or Handpiece Cord It could cause disconnection Be sure to attach NSK genuine tips when using NSK Varios ultras...

Страница 4: ... Control Plug Water Filter Case Water Filter Case Water Tube Water Tube Water Connector Water Connector 4 Handpiece Cord Connection Align the marks on the Control Unit and the Handpiece Cord Plug Gently push the Plug into the Connector CAUTION 1 Water System Setup Carefully check all Water Supply Connections prior to starting procedure Open the dental unit s water valve Set water pressure between ...

Страница 5: ...urface can be injured 2 Set the Power Level with the Power Level Key on the Front Panel The lamps in the Power Display Panel will indicate the selected power level Check to make sure the Power Level is set in the appropriate range for the selected Tip Fig 12 NOTICE Continuous pushing of the Power Level Key will cause the Power Level Display to climb When only water is desired without oscillating t...

Страница 6: ...lly by an object etc beeping starts to call attention To reset the alarm Release the Foot Control remove the object etc To enter new settings After confirming the settings of Water Mode Power Setting and Operating Mode push the memory key for approx 2 seconds The memory light will illuminate The old memory is cleared and new settings are programmed Memory Key Memory Lamp When the Tip is mounted to ...

Страница 7: ...p replacement is not necessary Yellow Wear of 1mm Tip is showing some wear Tip replacement is recommended Red Wear of 2mm Tip is badly worn Tip replacement is necessary Fig 20 OK NO Tip Card Handpiece Neck The Tip Card can be used to check the following tips G1 G4 G5 G6 G8 P1 P1D P10 and P20 Efficiency 1mm 2mm 25 50 Decrease Decrease 2 Changing Water Filter 1 Close the water valve of the dental uni...

Страница 8: ...oting No Poor vibration The tip is bent or broken The tip walks out Noise from the handpiece The handpiece is overheating The Power Lamp does not light even if the Power Switch is ON The tip does not generate vibration in spite of depressing the Foot Control The Power Cord or the Jack is disconnected The internal Fuse has burned out The tip is not tightened firmly Worn Tip Output has not been corre...

Страница 9: ... turns on and off Tip oscillates but light does not turn on Lamp is correctly and securely mounted in the socket but light does not turn on Power output is set at maximum at G mode Continuous operation for a long time at high mode G General Operation in a high temperature room Beeping while power on Beeping while stopping vibration of Tips The Water Adjustment Dial is closed Disconnected irrigatio...

Страница 10: ... measures may be necessary such as reorienting or relocating the Varios560 Varios560LUX b Over the frequency range 150kHz to 80MHz field strengths should be less than 3 V m Immunity test IEC60601 test level Electromagnetic environment guidance Conducted RF IEC61000 4 6 Radiated RF IEC61000 4 3 3Vrms 150 kHz to 80MHz 3V m 80MHz to 2 5 GHz Compliance level 3Vrms 3V m Portable and mobile RF communicat...