OPERATION MANUAL 

Please read this Operation Manual carefully  

before use, and file for future reference. 

Multi Function Ultrasonic Scaler 

OM-E0248E 004

’10.03.01 S

The EU directive 93/42/EEC was applied in the design and production of this medical device.

Содержание Varios 560

Страница 1: ...read this Operation Manual carefully before use and file for future reference Multi Function Ultrasonic Scaler OM E0248E 004 10 03 01 S The EU directive 93 42 EEC was applied in the design and production of this medical device ...

Страница 2: ...evice Safety instructions are classified as follows in accordance with the seriousness of the risk Use by medical professional such as doctor or dental hygienist is intended Do not unplug the Power Cord with wet hands to avoid electric shock Be sure to prevent water on the Control Unit because it may result in short circuit and electric shock Do not allow any impacton the Control Unit Do not drop T...

Страница 3: ...on and overheating outside the patient s oral cavity before use If any abnormalities are found stop using immediately and contact your dealer Should this products function abnormally cease operation immediately and returen it to the dealer for repair Do not force or pull on Power Cord and or Handpiece Cord It could cause disconnection Be sure to attach NSK genuine tips when using NSK Varios ultras...

Страница 4: ... Control Plug Water Filter Case Water Filter Case Water Tube Water Tube Water Connector Water Connector 4 Handpiece Cord Connection Align the marks on the Control Unit and the Handpiece Cord Plug Gently push the Plug into the Connector CAUTION 1 Water System Setup Carefully check all Water Supply Connections prior to starting procedure Open the dental unit s water valve Set water pressure between ...

Страница 5: ...urface can be injured 2 Set the Power Level with the Power Level Key on the Front Panel The lamps in the Power Display Panel will indicate the selected power level Check to make sure the Power Level is set in the appropriate range for the selected Tip Fig 12 NOTICE Continuous pushing of the Power Level Key will cause the Power Level Display to climb When only water is desired without oscillating t...

Страница 6: ...lly by an object etc beeping starts to call attention To reset the alarm Release the Foot Control remove the object etc To enter new settings After confirming the settings of Water Mode Power Setting and Operating Mode push the memory key for approx 2 seconds The memory light will illuminate The old memory is cleared and new settings are programmed Memory Key Memory Lamp When the Tip is mounted to ...

Страница 7: ...p replacement is not necessary Yellow Wear of 1mm Tip is showing some wear Tip replacement is recommended Red Wear of 2mm Tip is badly worn Tip replacement is necessary Fig 20 OK NO Tip Card Handpiece Neck The Tip Card can be used to check the following tips G1 G4 G5 G6 G8 P1 P1D P10 and P20 Efficiency 1mm 2mm 25 50 Decrease Decrease 2 Changing Water Filter 1 Close the water valve of the dental uni...

Страница 8: ...oting No Poor vibration The tip is bent or broken The tip walks out Noise from the handpiece The handpiece is overheating The Power Lamp does not light even if the Power Switch is ON The tip does not generate vibration in spite of depressing the Foot Control The Power Cord or the Jack is disconnected The internal Fuse has burned out The tip is not tightened firmly Worn Tip Output has not been corre...

Страница 9: ... turns on and off Tip oscillates but light does not turn on Lamp is correctly and securely mounted in the socket but light does not turn on Power output is set at maximum at G mode Continuous operation for a long time at high mode G General Operation in a high temperature room Beeping while power on Beeping while stopping vibration of Tips The Water Adjustment Dial is closed Disconnected irrigatio...

Страница 10: ... measures may be necessary such as reorienting or relocating the Varios560 Varios560LUX b Over the frequency range 150kHz to 80MHz field strengths should be less than 3 V m Immunity test IEC60601 test level Electromagnetic environment guidance Conducted RF IEC61000 4 6 Radiated RF IEC61000 4 3 3Vrms 150 kHz to 80MHz 3V m 80MHz to 2 5 GHz Compliance level 3Vrms 3V m Portable and mobile RF communicat...