Taper Ring
15
FR
EN
Fig. 14
· Do not autoclave the handpiece with other instruments even in a pouch. This is to prevent possible discoloration and damage to the handpiece
from chemical residue on other instruments.
· Keep the handpiece in normal atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
· Immediately after use the handpiece should be cleaned, lubricated and sterilized. If blood remains on any external or internal surfaces it can
harden and cause rust.
· Do not heat or cool the handpiece too quickly. Rapid change in temperature could cause damage to the handpiece.
· Do not use the dry cycle if the sterilizer chamber temperature may exceed 135˚C.
· Autoclave sterilization is recommended for this product. The validity of other sterilization methods is unknown.
· Do not touch the handpiece immediately after autoclaving as it will be very hot.
NOTICE
NSK recommends Class B sterilizers as stated in EN13060.
8. Replacing the O-rings (X450L / X450)
Replace the O-rings if water is present in the exhaust air line. ALWAYS change the complete set of O-rings.
1) Loosen and Remove the taper ring at the rear of the handpiece. (Fig. 14)