(4) "Pd": Knee Controller or Foot pedal check

Press      Forward/Reverse Selector Switch, and      Indicator will change. During normal time, 
Indicator changes in hexadecimals ( 0~9, A~F ) according to the amount of depressing 
     Knee controller or      Pedal. Also, depressing the pedal slightly lights LED, and depressing 
it fully extinguishes the lamp. If      Indicator does not change smoothly or LED does not light 
properly,      Knee controller or      Pedal may be faulty. To cancel this check, press 
     Forward/Reverse Selector Switch again.

(5) "in": Initializing function

Press      Forward/Reverse Selector Switch, and beeps will be made and rotation direction, 
other settings will return to the factory set condition.

Rotation direction: FWD (forward)
Vacuum-coupled mode: OFF
Motor/Turbine: Motor
AUTO Cruse: OFF

(3) Motor signal check 

(Hall IC check)

Press      Forward/Reverse Selector Switch, and      Indicator will indicate one or two 
horizontal lines. Turn the motor slowly by hand, and this display will change to one line, two 
lines, one line, two lines, smoothly from the top to bottom or from the bottom to top. If any 
one of the three lines does not light, the sensor (Hall IC) in the motor is faulty or the cord is 
severed, therefore repair is needed. To cancel this check, press      Forward/Reverse Selector 
Switch again.

Eg.

The unit is provided with a maintenance mode to check the functionality of the switches, 
display, foot pedal, motor, etc. While pressing      Turbine/Motor Selector Switch and      Auto 
Cruise Switch at the same time, turn on the Power Switch and keep pressing the button until 
beeps are made (for about 2 seconds). At this time, turning the volume from the minimum 
position in order displays "oP", "dP", "HL", "Pd" and "in", allowing the following checks. To 
release Maintenance Mode, turn Power Switch off and switch on again.

(1) "oP": Switch check 

(operation check)

Press the switches on the panel, and the right and/or left lamps will light to check to see if 
the switches operate normally.

(2) "dP": Display check

Press      Forward/Reverse Selector Switch, and the lamps will light one by one to check to 
see if they operate normally. To cancel this check, press      Forward/Reverse Selector Switch 
again.

10

11

9

9

8

10

15

15

15

15

Fig.8

Fuse is located in     Fuse Box. Release ratchet 
clamp located on the top and bottom of the 
     Fuse Box and pull it out to change the fuse 
(T3.15AL250V). (Fig. 8)
When the fuse come down, Insert new fuses 
(T3.15AL250V)into the      Fuse box and then 
push the      Fuse box into the original position 
until hearing a sound click.

- Fuse: Order No. (120V) : 12001-31510
- Fuse: Order No. (230V) : 12001-16010

9. Replacement of Fuse

10. Maintenance Mode

On some dental tables with vacuum dust collector, the motor may be used while being coupled 
with a dust collector. When such a dust collector * is used, power consumption of ULTIMATE 
XL-K/D can be regulated so that the vacuum-coupled function can work. If you need coupling 
with a vacuum dust collector, select the mode as follows:

How to select the mode

While pressing      Forward/Reverse Selector Switch, turn on      Power Switch, and the mode 
can be selected. A long beep indicates vacuum-coupled mode and 2 short beeps indicate 
non-coupled (energy-saving) mode.

* Each time the switch selection is made, the mode changes between vacuum-coupled mode 

and non-coupled mode.

* A currently known dust collector is KAVO EWL-560.

11. Vacuum-coupled Mode

CAUTION

Fuse is burned out when a short circuit occurs or when over-voltage is flowed into the 
primary current source. If the cause is uncertain, return the product to an authorized 
NSK’s service shop for inspection.

9

9

6

6

9

5

5

9

9

9

11

11

12

11

2   3   4      A   b   C   d   C   b   A   4   3   2

English

Содержание NE231

Страница 1: ...OPERATION MANUAL Micromotor for Laboratory Work English ...

Страница 2: ...ine Option 6 Protective Circuit 7 Memory Function 8 Error Code 9 Replacement of Fuse 10 Maintenance Mode 11 Vacuum coupled Mode 12 Handling of Motor and Handpiece 13 Handpiece Stand 14 Specification 15 Troubleshooting 16 Disposing Product 1 5 6 7 8 8 8 9 9 10 10 11 12 14 14 15 18 Contents WARNING When using electric tools basic safety precautions should always be followed to reduce the risk of fir...

Страница 3: ...zed cord will cause a drop in line voltage resulting in loss of power and overheating Table below shows the correct size to use depending on cord length and nameplate ampere rating If in doubt use the next heavier gauge The smaller the gauge number the heavier the cord 9 Install equipment so that the power supply cord can be pulled out without hindrance in event of emergency 1 For your own safety ...

Страница 4: ...oto may cause damage 25 Do not turn the bur lock Chuck Control Ring while the handpiece is rotating 26 Do not rotate the motor when the bur lock Chuck Control Ring is at OPEN position or a bur is not mounted in the chuck It may cause the motor handpiece disconnection or sudden heat generation 27 Activation of Circuit Breaker means too much load is applied to the motor beyond the capacity the motor...

Страница 5: ...t to the switch will light up If the desired speed is set with the Knee Controller or the Pedal and a uniform speed can be maintained for one second the motor will continue running at the desired speed if the Knee Controller or the Pedal is released To cancel it depress the Knee Controller or the Pedal again or press Auto Cruise Switch again 6 7 1 Mounting of the Unit 1 ULTIMATE XL K Bore a hole w...

Страница 6: ... given to the rotation direction 7 Memory Function When the motor is stopped due to some trouble such as a failure overload wire breakage or misuse Speed Display displays the error code for checking the status of the unit and understanding the cause of the trouble easily 8 Error Code Fig 7 CAUTION Use of 40 000min 1 rpm or more is allowable only when the bur manufacturer or dealer specifies that t...

Страница 7: ... To release Maintenance Mode turn Power Switch off and switch on again 1 oP Switch check operation check Press the switches on the panel and the right and or left lamps will light to check to see if the switches operate normally 2 dP Display check Press Forward Reverse Selector Switch and the lamps will light one by one to check to see if they operate normally To cancel this check press Forward Re...

Страница 8: ...tc accumulate in the chuck and the bur is caught insecurely causing runout OPEN LOCK Loosen Loosen Loosen Fig 11 Loosen Loosen Loosen 3 Insertion of Chuck Thinly apply oil before insertion Open the Chuck Control Ring insert the dummy bur or the bur in use into the chuck and turn the chuck clockwise by hand until it stops Then lock the Chuck Control Ring and the chuck could hold the bur securely Fi...

Страница 9: ... cord and check the operation The motor cord is disconnected If you have two or more units replace the motor and the motor cord and check the operation Insert the power plug correctly Replace it with a specified fuse If the reason the fuse has blown is unknown ask for an inspection Ask for repair Power on the Control Unit without pressing the Foot Pedal When it does not work definitely take off th...

Страница 10: ...unit Ask for repair of the turbine adapter and or the Ultimate XL K D unit The motor and handpiece do not run or Reset Lamp lights Error code E3 appears Error code E4 appears Error code E5 appears Error code E6 appears Error code E8 appears The motor cord is disconnected Maintenance mode 3 HL Confirm whether you work by a motor signal check normally The chuck is open Check turning a tip with a fin...

Страница 11: ...18 Please consult with dealer from whom you purchased it about waste disposal 16 Disposing Product ...

Отзывы: