REC-1067 Rev B
13
Notice:
The Speed and Time CANNOT be adjusted while the instrument is running.
Starting a Run
To start the run, depress the START/STOP. The START/STOP led will illuminate when the
instrument is running. Once run is complete, clear the completed run by depressing any arrow
key.
Stopping a Run
To stop a run at any point, depress the START/STOP key. The device will slow to a stop and the
instrument will return to the home position. The run parameters will be saved at the stopping
point. These parameters can be adjusted if needed.
When ready to restart a run, depress the START/STOP key.
Opening the Door During a Run
Opening the door during a run will immediately turn off the motor and stop the run. If the door
is opened mid-run, follow the instructions on the screen to restart the run from the stopping
point:
•
Depress any arrow key
•
Depress the START/STOP key to home the instrument
The run parameters will be saved at the stopping point. These parameters can be adjusted if
needed.
When ready to restart the run, depress the START/STOP key.
Useful Suggestions
Loading and Unloading the Rack:
Always use two hands (one on each handle) to load/unload the rack onto the rack mount pins.
Ensure the rack is fully engaged with all three rack mount pins before starting a run.
Loading and Unloading the Sample Cartridges:
Insert the cartridges straight down into their respective cartridge slots to load. Remove by
pulling straight up out of the cartridge slot. Do not manually manipulate the spring clips while
doing this as aggressive manipulation of the spring clip may damage it. The spring clip will be
moved aside without additional manipulation as you push or pull the cartridge in and out of the
slot.
Always use two hands to load/unload the cartridges into the rack. One hand should hold on to
the rack while loading/unloading the cartridges.
Confirm that the cartridges are completely set in the recessed area of the cartridge slot to
ensure that the rack clips are fully engaged.