REC-1067 Rev B
12
•
Digital display
•
Power indicator light
•
Stop/Start button and indicator light
•
Time adjustment keys
•
Speed adjustment keys
Chapter 3 - Operating Instructions
1.
Turn on the power switch located in the back of the instrument. The POWER light on the
user interface will illuminate.
2.
Ensure that the instrument is in the home position. If the instrument is not in the home
position, the display will read “press start to home”. Press the start key to home the
instrument.
3.
Open the mixing chamber door.
4.
Place the Revolution Cartridge Rack containing up to 8 Revolution nRicher Cartridges
into the instrument.
5.
Load the Revolution Cartridge Rack onto the rack mounting pins by inserting top of
cartridge rack (single hole) onto the top pin and lining up the bottom pins with the
bottom slots in the rack. Slide the rack under the flange feature of the pins. Ensure that
the rack is captured by all 3 pins.
6.
Close the mixing chamber door. The door must be fully closed to operate the
instrument.
7.
Use the arrow keys to adjust the speed and time according to the appropriate
Revolution Isolation Kit program located in the Revolution Isolation kit Instructions for
Use. The timer range is between 1 min and 999 minutes and the speed range is between
10 RPM and 30 RPM.