Maintenance
The only maintenance required is the annual replacement of the Ionizer to ensure maximum
effectiveness of the ionizer. It is required to return the expired device to NRD for proper
disposal.
Calibration is not required.
3.0 Trouble Shooting Guide
Symptom
Probable Cause
Corrective Action
1.
Static charges are not
eliminated
Ion cell may be out of date and
need replacement.
Possible oil contamination of ion
cell.
Ions generated are being
recombined before they reach the
charged surface.
Call NRD for Ion Cell
replacement.
Return cell to NRD for
evaluation. Do not attempt to
clean the device.
Position the fan closer to the
charged part.
2.
Fan does not turn on
No power.
Blown fuse internal to fan.
Check outlet and breakers.
Check for visible damage to
power cord. Return fan to NRD
for evaluation.