1.

 

Ensure the AB-ABH Ionizing Head(s) are connected to the AB-TC Controller according to guidelines in 
the installation section. 

2.

 

Plug the controller into a properly grounded outlet and turn power switch, located on the back of the unit, to 
“ON”. 

3.

 

The green LED on the front of the AB-TC Controller should be on. If the red LED is on, or no LEDs are 
on, this indicates a problem.  Check for loose connections and if problem continues contact NRD, LLC for 
assistance. 

 

2U500 Staticmaster Ion Source 

To maintain peak performance it is recommended to replace the 2U500 Staticmaster annually.  Notification will be 
given prior to the end of each year. Upon arrival of a replacement, change out the 2U500 Staticmaster and send the 
old 2U500 to NRD LLC. 
 

Cleaning 

Use only a diluted IPA solution and a cleanroom-compatible cloth to clean any accumulated dirt on the controller or 
ionizing head(s).  Change cloth frequently to ensure that the dirt is completely removed.  Do not use any solvents 
that may damage the powder coat finish of the controller. 

 

IEC Indications 

High Voltage – the rear panel output of the AB-TC Controller is via an MHV connector.  The voltage on this 
connector can be as high as ±2400V.  Note that the source impedance is in excess of 1MW so no harm or discomfort 
will come to the operator by touching the contact. 
The voltage source on the board is much lower impedance and, as such, contact with high voltage section of the 
circuit board would be painful.  This instrument uses high voltages that can cause personal injury.  Before servicing, 
shut down the instrument and disconnect the instrument from line power.  It is recommended that the AB-TC 
Controller not be opened. 

 

The MHV connector on the chasis and the mating cable connector are both rated at 5000V.  The RG 59 cable used 
to connect the AB-TC Controller to the AB-ABH Ionizing Head is rated at a working voltage of 2400V. 
Rear Panel Auxiliary Connector signals do not exceed 10V and require only insulation of connections to protect 
against 10V. 

 

Note: if the equipment is used in a manner not specified by the manufacturer, the protection provided by the 
equipment may be impaired. 

 

 

- This symbol means that there is high voltage present at this connector. 

 

- This symbol means that it is necessary to connect this binding post to ground for proper operation of this 
  instrument. 

 

- This symbol means that in order to have safe and proper operation of the device, the user should read the 
  manual. 

 

Specifications: 

Ionization Source: 

500 µCi of 

210

Po. (provides full ionization performance for 12 months) 

Boost:   

 

±2400 V, Output from Model AB-TC via MHV connector. 

Operating Period:  

500 ms 

Balance Voltage:  

Factory preset 

Operating Range:  

AB-ABH Ionizing Head, 0° C to 80° C 

 

 

 

AB-TC Controller, 5° C to 50° C 

Power Requirement: 

24 V DC, <200 mA (provided by 100 - 240 V inline power supply) 

Offset Voltage:   

Factory preset 

Alarm Condition:  

Positive or Negative Boost out of regulation by 5% or actual output ceases to 
cycle. 

Alarm Output:   

User Selectable either contact closure or pull up voltage to +5 V (user option). 

Содержание Alphaboost AB-TC-1

Страница 1: ...on Maintenance Staticmaster Alphaboost Model AB TC 1 AB TC 1B Contents Alphaboost System General Description Page 2 Model AB TC Control System Page 2 Model AB ABH Ionizing Head Page 3 Model 2U500 Staticmaster Ionizer Page 3 Model AB TC 1B Alphie Page 3 Installation Page 4 AB ABH Ionizing Head Installation Page 3 2U500 Staticmaster Installation Page 4 AB TC Controller Installation Page 4 Operation ...

Страница 2: ...ntroller The AB TC controller see figure below is used to create the waveform to drive the ions away from the ionizing head It is factory preset for ON time for each polarity of the Boost Voltage The amplitude of the ionizer Boost Voltage The Duty Cycle of the Boost Waveform The controller includes a readout of the duty cycle Balance setting of the controller with a resolution of 1 While the setti...

Страница 3: ...put of the AB TC Controller connects up to 4 ionizing heads when using the Splitter model AB SPL sold separately AB ABH Ionizing Head AB SPL 1 to 4 Splitter Model 2U500 Staticmaster Ionizer The model 2U500 Staticmaster is a small self powered ion producing device which contains 500 µCi of 210Po All NRD ion sources are encapsulated in precious metals by means of a special patented process This resu...

Страница 4: ...op panel of the Analytical Balance Draft Chamber Center metal electrode as close as possible over the weighing pan 3 Attach BF3 EZ Mount to inside of top panel of the Analytical Balance Center under metal electrode on AB ABH ionizing head 2U500 Staticmaster Installation The 2U500 Staticmaster slides into the BF3 EZ Mount that is attached inside the Analytical Balance AB TC Controller Installation ...

Страница 5: ... will come to the operator by touching the contact The voltage source on the board is much lower impedance and as such contact with high voltage section of the circuit board would be painful This instrument uses high voltages that can cause personal injury Before servicing shut down the instrument and disconnect the instrument from line power It is recommended that the AB TC Controller not be open...

Страница 6: ...etion of NRD free of charge Any device under warranty should be shipped prepaid to the NRD factory Call Customer Service at 716 773 7634 for a Return Authorization number and shipping instructions Include a copy of your original packing slip invoice or other proof of purchase date If the device is out of warranty NRD LLC will quote repair charges necessary to bring your device up to factory standa...

Отзывы: