14
Integral needle
(for PMU)
Connecting the Roller/needle
Use disposable gloves
Make sure the ORON 58 control unit is powered off
Take the roller/needle out of the sterile packaging (make sure the package is not
damaged prior to opening it, if it is, do not use it)
Insert roller/needle into the handle roller base
Rotate the roller/needle a quarter turn clockwise to lock it in place. Make sure the
roller/needle is properly locked by trying to pull it gently out. If needed repeat the
roller/needle insertion.
Follow section 9 (Roller adjustments)
Activate the handle by pressing the foot pedal
Disconnecting the Roller/Needle
Make sure the ORON 58 control unit is powered off
Turn the roller/needle a quarter turn counter clockwise and pull it out of the roller
base
Dispose of the roller/needle properly
Place the handle back into the control unit
Warning:
The ORON 58 device can only be used by NPM authorized and trained staff
Needle
Head
Insert area
Locking Mechanism