Functional Description (Base NR2310D-ROG)
Outputs
Each output is fault and electrostatic discharge protected. Each output is
independent and any output can be faulted for an indefinite period of time with
no permanent damage. Each output is connected to a monitor circuit that
detects a local fault on the output. The fault status is indicated on the front
panel via an LED or reported in the digital display. The fault status and the
protection on each output facilitates installation. A channel fault will not
activate an “ALERT” state and the status relay will not be opened.
The standard outputs are1.0 Vrms 10 MHz sine into 50 Ohm.
Built-in Test
The built in test monitors the following:
Power Supplies - All power supplies are monitored. If a supply fails to meet test
limits, an alert is generated.
Channel Faults - if a channel fault is detected, an indication is given but an
alert is not generated.