Latch Channel Value
The latch channel values allows the user to save the current channel output
value for use as the reference value for alert settings.
A channel alert is triggered when the channel output voltage exceeds or falls
below a percentage of the reference value. This reference value is 1.10Vrms as
a default, but can be set by the user.
There are two ways to set the reference voltage. The RS232 serial port allows
for setting an individual channel’s reference voltage with the $SET command.
The user can also use the latch channel values to take a snapshot of all current
outputs and use these as the reference values.
Save Configuration
The save configuration screen allows the user to save the current settings for
alert threshold, input threshold, attenuation, input select, reference voltage and
any other settings that have been modified via the RS232 port.
The save configuration action is equivalent to the $SAVEFL command on the
serial port.