Novis ProCair Quad Скачать руководство пользователя страница 1

1

ProCair Quad 

User Guide

 

U S E R   G U I D E

ProCair Quad

Содержание ProCair Quad

Страница 1: ...1 ProCairQuad User Guide U S E R G U I D E ProCairQuad...

Страница 2: ...provement and reserves the right to amend specifications presented in this guide Information correct at time of production February 2019 2019 Novis Healthcare All rights reserved Important Notice Defi...

Страница 3: ...utions 6 7 System Preparation 8 10 Patient Set Up 11 Operation Control Unit 12 13 Operation Mode 14 Operation Pull Cap CPR Release 14 Operation Transport Mode 15 Operation Deflation Storage 15 Care an...

Страница 4: ...ial and venous capillary occlusion in the patient s surface tissue maintaining and stimulating the flow of blood and lymphatic fluids through these tissues to provide essential oxygen and remove metab...

Страница 5: ...vironment Temperature 5 C to 60 C 41 F to 140 F Humidity 30 to 90 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The ProCair Quad mattress...

Страница 6: ...part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams objects in pockets etc Never use sharp objects or electrical...

Страница 7: ...e of smoking materials or open flame air flowing through the mattress will support combustion Do not open the control unit risk of electrical shock Refer servicing to qualified service personnel Dispo...

Страница 8: ...ase Ensure the buckles are securely fastened and straps are pulled tight Do not secure mattress straps to bed side rails straps will tear Ensure that straps do not interfere with the operation of the...

Страница 9: ...ith other control unit which looks similar Straighten any twists in the umbilical cord air hoses to ensure uninterrupted air flow between the control unit and mattress Ensure the umbilical cord is not...

Страница 10: ...alarmguideonleft ALARMINDICATORS BATT IN USE 9When initial inflation is complete the Pressure Setting and Max Inflate indicators will extinguish to indicate the system is ready for use The system auto...

Страница 11: ...the cover to feel the cells for inflation 2With the patient lying supine unzip one side of the top cover just past sacral region lower spine 3Slide your hand underneath the patient and feel for a def...

Страница 12: ...de on the control panel to look up the fault indicated also refer to the corresponding troubleshooting guide page 19 for detail Alarms include LOW PRESSURE possible loss of pressure HIGH PRESSURE matt...

Страница 13: ...ut not switched on When periodical servicing is required the amber Standby light will flash Please contact Novis Healthcare for system maintenance This equipment must only be serviced by a qualified s...

Страница 14: ...g the risk of bottoming out Once the head of the mattress is brought back to a reclined position or below a 5 angle Fowler Boost mode will automatically disengage FOWLER ALTERNATE STATIC Backup Batter...

Страница 15: ...sed Heel Therapy The specialised HeelCair zone enables the seven narrow alternating air cells in the heel zone to be disconnected to remove all pressure on a patient s heels without interruption to ma...

Страница 16: ...T IN USE Deflation and Storage 1 Press the power button for a minimum of three seconds to switch off the control unit 2Switch off mains power and unplug the power cord from the mains outlet 3Open CPR...

Страница 17: ...se and Cells Cleaning Swab the mattress base and cells with a solution of sodium hypochlorite or similar up to 10 000 ppm available chlorine Dry thoroughly before reassembly Do not machine wash or tum...

Страница 18: ...the zipper running towards the CPR Then close both zippers by zipping towards the foot end of the mattress Air Filter Replacement 1 Switch off power supply to the control unit and disconnect the air h...

Страница 19: ...power after 1 minute and restart initiation Remove patient from mattress Turn off control unit and disconnect process return patient to mattress once initiated and alternation has resumed If issue per...

Страница 20: ...power supply Check for loose power cord connection and ensure main power is switched on Check the fuses in control unit Replace if necessary Checkconditionofpowercordandplug Checkifmainssocketisfault...

Страница 21: ...T 0 2 A MAX FUSE RATING T1AH 250V MIN MAX PRESSURE 20 60 mmHg 6 mmHg PROTECTION TYPE Class II Type BF INGRESS PROTECTION RATING IP21 MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEI...

Страница 22: ...onsible and contact your local authority on available options to recycle this product at its end of life Service Life The expected service life of a control unit and a mattress is highly dependent on...

Страница 23: ...t a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum ou...

Страница 24: ...60 Hz 50 Hz Conducted RF IEC 61000 4 6 3 Vrms 0 15 MHz 80 MHz 3 Vrms 0 15 MHz 80 MHz Portable and mobile RF communications equipment should be used no closer to any part of the ProCair Quad control u...

Страница 25: ...Hz 0 2 0 3 9 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation B 18 Hz 2 0 3 28 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25...

Страница 26: ...may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for any repair other than th...

Страница 27: ...27 ProCairQuad User Guide Notes...

Страница 28: ...Pressure care and patient handling specialists novis com au 1300 738 885 V1 June 2019...