11
Lifetime
Reprocessing Restrictions, page 21 ]
12
Shelf Life and Storage
Before first processing:
Store the product in unopened original packaging.
13
Possible Complications and Side Effects
As with underlying disease and procedure.
14
Combining with Other Procedures
Does not apply.
15
Reprocessing
WARNING
•
The product is not sterile. Process the product before first and any further
application.
This is the only way to ensure the product is germ-free and functional.
Only use the disinfection and sterilisation procedures validated for the device and
product.
Ensure regular inspection and maintenance of the devices in use.
Adhere to the validated parameters in every cycle.
Observe the local regulations and hygienic rules of the medical office or hospital.
15.1
Reprocessing Restrictions
The product is suitable for multiple reprocessing.
The following factors are relevant for determining the product's lifetime:
• Reprocessing procedure
• Wear and tear
• Chemicals used
Do not use damaged / worn products.
15.2
Cleaning and Disinfection Agents
Do not use cleaning and disinfection agents that contain the following substances:
• Organic, mineral or oxidising acids
• Strong lyes
• Organic solvents (e.g., alcohols, ethers, ketones, benzines)
• Halogens (chlorine, iodine, bromine)
• Aromatic / halogenised hydrocarbons
• Heavy metal salts
• Aldehydes
• Scratching substances
English