6
If the RS232 connection is sound, the NovaLynx Corporation header and the software
version will appear on the screen.
WeatherPort WS-16-A
Modular Weather Station
NovaLynx Corporation Inc.
Version: 4.0.8 (c)1996-2006 Mesotech
Protected by USA and Int'l copyright laws
Press “m”, “m” for main menu
Figure 1
If nothing appears on the screen, unplug the serial cable from the WS-16 module and
plug in the WS-16 RS232 Output Tester supplied with the unit. If the module is
functioning, the red light should flash every 5 seconds.
If the module is functioning but you still see nothing on the screen, the wrong serial port
may be selected. Try selecting Com 2 or another functioning serial port. You can see a
list of available ports in My Computer > System Properties > Device Manager > Ports.
2.4 Sensor Installation
Install the sensors in their chosen locations, bearing in mind the installation
considerations noted earlier. Run cables from the sensors to the data acquisition
module location, with no cable exceeding the maximum allowable length listed below.
Sensor
Maximum Cable Length
Wind*
250' (76m)
T/RH*
250' (76m)
Rain*
900' (275m)
Barometric Pressure**
10' (3m)
Solar Radiation*
100' (33m)
*Std cable length 40' **Std cable length 18"
When the sensors have been installed and the cables run, connect and test them as
described in the following sections. Refer to the hook-up drawing at the back of this
manual to verify wire colors and physical connections.