19
Running Ketone Control Solution
Important Information for Ketone Control Solution
• Use only Nova Max Plus Control Solutions for the test.
• Check the expiration date on the control solution vial. Do
not use control solution past the expiration date or you may
get inaccurate results.
• Store only for 3 months after first opening. When you open
a new vial of control solution, count forward 3 months and
write that date on the label of the control solution vial. Dis-
card any remaining solution after the date you have written
on the vial.
• Store the control solution tightly closed at room tempera-
ture below 86°F (30°C). Do not refrigerate or freeze.
• Shake the control solution well before using.
Nova Vet Monitor OG PN52899E EN.indd 19
4/14/15 10:26 AM