Setting the Ketone Slope
Set the Slope for Ketone:
Press the Left/Right keys to decrease/increase the Slope value
in increments of 0.01. The default value is 1.00 and the range is
0.50 to 1.50.
IMPORTANT:
If
necessary, contact
Technical Support for
guidance on changing
the ketone slope value
from the 1.00 default
setting. This change
will modify the reported ketone result for whole blood specimens
only. Quality Control specimens, designated as a control, are not
affected by this change.
Lowest Default Highest
Slope Slope Slope
Nova Vet Monitor OG PN52899E EN.indd 15
4/14/15 10:26 AM