53
Codes
Code Alerts
Action
E-0
Software error detected
Perform the test again. If you get the
same error, call Nova Technical Support.
E-1
Hardware error detected
Perform the test again. If you get the
same error, call Nova Technical Support.
E-2
Meter temperature is outside
Move the meter to an area
of the range for testing
where the temperature is acceptable
(41°-113°F), allow meter to adjust to the
temperature. Repeat testing.
E-3
Defective or previously used
Repeat test with a new
test strip is detected
test strip.
E-4
A short sample (control
Repeat test with a new test
solution or blood) is identified strip and adequate sample.
LO or If test results are outside of the
Repeat test. If you get the
HI
measuring range of the meter
same result, consult your physician.
Battery level is low
Replace the battery
Error Messages