Testing a Quality Control Solution
1. Grasp a test strip with logo side up and gold
side down. Then insert the gold end into the
meter. (See Meter figure.)
2. Identify the sample as a control; Use the
left
or right
arrow buttons
to find the de
-
sired control level: QC1
or QC2
.
3. Shake the control solution well before each use. Discard
the first drop.
4. Place a drop of control solution from the
bottle at the end of the test strip until the
solution is drawn into the well of the test strip.
5. A Lactate quality control test result is avail-
able on-screen in 13 seconds.
6. Remove the strip manually or use the ejector button on
Testing: Blood and Quality Control Solutions