6.2 Mikromotor Endo
1) The micromotor, cable and plug should be cleaned after each use to prevent the build up of
deposits and debris that can destroy it, if not removed. Wipe it with a clean cloth dampened with
disinfectant solution.
2) Put the sterilizable stopple on the motor.
3) After packing motor and cable, autoclave it at maximum 134°C .
4) After autoclaving, allow motor to dry for one hour at room temperature.
Store motor bagged and tagged with sterilization date!
•
Do not bend the motor cable!
•
Clean, disinfect and sterilize the Micromotor after each use
•
Do not clean Micromotor with compressed air
7 Troubleshooting
Problem:
Cause:
Solution:
Device does not operate at
all:
Plug is not inserted properly
Insert plug and check fitting
Wrong operating voltage
Check mains voltage output
→
please see operating voltage on the
rating plate
Footswitch fails to function:
Pedal not connected
Plug pedal cable into rear of control
unit
Control unit not turned on
Turn main switch ON/OFF
Micromotor does not run:
Motor not turned on
Turn motor with „Motor“-key on or with
footswitch
Motor not connected
Connect motor cable to control unit
Motor incorrectly attached
Press handpiece firm on motor until
handpiece snaps in
In the event a problem cannot be solved, contact dealer or authorized service technicians, as listed on
the last page of your Operating Instruction Manual.