5.2  The footswitch 

The TCM Endo III can be operated with or without footswitch. Press footswitch to start the motor. 

Footswitch not pressed:..................Motor off 
Footswitch fully pressed: ................Motor on 

The tool speed corresponds to the speed adjusted on the control unit. 

5.3  Automatic Limiter (AL) 

The Automatic Limiter (AL) limits the torque level applied to the instrument. If the selected torque level is 
reached, the speed will drop to zero rpm. A torque level equal to the selected setting will still be exerted 
against the file. 
All torque values are shown in Nmm (for example a setting of AL 20

 indicates that the instrument can 

be loaded to maximum of  20 Nmm before rotations stops). 

Adjusting the AL mode: 

1)  Press the key “ATC SET”. In the display appears the torque mode set  (for example [AL   20]). 

2)  Pressing the keys “

” and“

” will increase and decrease the torque level respectively. 

3)  By pressing the  “ATC SET” briefly once more the digital display switches back automatically. 

5.4  Automatic Protection (AP) 

The automatic protection (AP) mode is a special function against file breakage. 
When  the  selected  torque  level  is  reached,  the  motor  reverses  immediately  to  free  the  file  and  then 
returns automatically to forward operation. 
All torque values are shown in Nmm (for example a setting of AL 20

 indicates that the instrument can 

be loaded to maximum of  20 Nmm before rotations stops). 

 

 

 

The Automatic Limiter (AL) can be adjusted in the following ranges: 

 

1:1 

 from 1 to 19 Nmm 

 

8:1 

 from 2 to 50 Nmm 

 

16:1 

 from 5 to 100 Nmm 

 

If the display says  [AP.....] : 
press the key “

” to get back to  [AL.....]. 

The  green  light  will  illuminate  when  the  AP  -  mode  is  activated,  the  green  light 
does not illuminate on the AL- mode. 

 

The automatic protection AP can be adjusted in the following ranges: 

 

1:1 

 from 1 to 19 Nmm 

 

8:1 

 from 2 to 50 Nmm 

 

16:1 

 from 5 to 100 Nmm 

Содержание TCM Endo III

Страница 1: ...NOUVAG Made in Switzerland Nr 31817 www nouvag com Operation manual GB TCM Endo III ...

Страница 2: ...d Storage Environment 2 1 5 Description on device 2 2 Safety measures 3 3 Shipping contents 3 4 Set up 3 5 Operation 4 5 1 The TCM Endo III unit 4 5 2 The footswitch 5 5 3 Automatic Limiter AL 5 5 4 Automatic Protection AP 5 5 5 Memory 6 6 Disinfection Cleaning and sterilization 6 6 1 Control unit and footswitch 6 6 2 Mikromotor Endo 7 7 Troubleshooting 7 8 Disposal 7 ...

Страница 3: ...l preparation With use of the AP mode file stress is greatly reduced 1 2 Technical Data Voltage 115V 50 60Hz or 230V 50 60Hz see device label Fuses 2 x T2A for 115V and 2 x T1A for 230V Power 40 VA Type Application part type B Speed 1 200 16 000 rpm Dimension 90 x 100 x 170 mm Net weight 2 kg 1 3 Operating Environment Relative Humidity max 80 Temperature 10 to 40 C Barometric Pressure 800 to 1060h...

Страница 4: ... nor can Nouvag AG be held liable for any injuries to persons or animals in any case when the device is miss used or not operated applied or maintained in strict accordance with the user owner instructions set out in the operating manual In the event of any doubt or question the user is to contact Nouvag AG or its lawful representative for clarification or assistance 3 Shipping contents 1520 Contr...

Страница 5: ...r reduce torque Rev Forw To change motor direction of rotation a warning tone will sound when the motor operates in reverse mode 1 1 Press this key if using handpiece 1 1 Speed from 1 200 to 16 000 rpm Torque 1 2 3 4 5 7 10 13 16 19 Nmm 8 1 Press this key if using contra angle 8 1 Speed from 150 to 2 000 rpm Torque 2 4 6 10 15 20 25 30 40 50 Nmm 16 1 Press this key if using contra angle 16 1 Speed...

Страница 6: ...20 2 Pressing the keys and will increase and decrease the torque level respectively 3 By pressing the ATC SET briefly once more the digital display switches back automatically 5 4 Automatic Protection AP The automatic protection AP mode is a special function against file breakage When the selected torque level is reached the motor reverses immediately to free the file and then returns automaticall...

Страница 7: ...olving agents for cleaning Autoclave material in transparent packaging Do not fill the sterilization bag more than 80 Autoclave material at 134 C maximum Sterilized material should be stored and tagged with sterilization date 6 1 Control unit and footswitch The control unit and the pedal do not come into contact with patients Only clean the surface of the unit do not use harsh cleaners or solvents...

Страница 8: ...properly Insert plug and check fitting Wrong operating voltage Check mains voltage output please see operating voltage on the rating plate Footswitch fails to function Pedal not connected Plug pedal cable into rear of control unit Control unit not turned on Turn main switch ON OFF Micromotor does not run Motor not turned on Turn motor with Motor key on or with footswitch Motor not connected Connec...

Отзывы: