ES3 "Intelligent" Pressure and Level Transmitter
EX-1861
• The transmitter will return to the home screen.
• Choosing Reset will put the transmitter back to factory setting (0.0 seconds).
READOUT
In this menu, the type of readout on the display can be adjusted.
Factory Setting = Unit
1. Navigate to program point
P109
, and push the menu button to enter the menu.
2. Four choices appear on the screen:
Current
: Current value (4 - 20 mA)
Unit
: Pressure unit (Selected in
P104
)
Percentage
: 0 - 100%
Temperature
: Actual process temperature (°F or °C) Indication , accuracy depending on
sensor position.
3. Navigate to the desired choice, confirm by pushing the menu button.
4. The transmitter will return to the home screen.
BURST MODE (HART
®
)
The transmitter (Only when HART
®
is present) can be configured for Burst mode.
This will enable continuously broadcasting standard HART® reply messages.
1. Navigate to program point
P110
, and push the menu button to enter the menu.
2. Five choices appear on the screen:
Mode Cntrl, Cmd number, Message, Period
and
Trigger
3. Select
Mode Cntrl,
and push to confirm.
4. Two choices appear on the screen:
On
and
Off
•
Choose
On
to turn on burst mode.
•
Choose
Off
to turn off burst mode.
5. Select
Message
to select the burst message 0,1,2 or 3 and push the menu button to confirm.
6. Select
Cmd number,
and push the menu button to confirm.
Five choices appear on the screen:
•
Cmd 01
= PRIMARY VARIABLE
•
Cmd 02
= CURRENT AND PERCENT OF RANGE
•
Cmd 03
= DYNAMIC VARIABLES AND CURRENT
•
Cmd 09
= DEVICE VARIABLES WITH STATUS
•
Cmd 48
= ADDITIONAL TRANSMITTER STATUS
Choose the preferable burst mode, and push menu button to confirm.
7. Select
Period
, and push the menu button to confirm.
Two choices appear on the screen:
Max Time
and
Min Time
•
Select
Max Time
to set the maximum amount of time when the message will be
send. This value can be set from 0.5 to 3600 seconds.
•
Select
Min Time
to set the minimum amount of time when the message will be
send. This value can be set from 0.5 to 3600 seconds.
Enter the preferred value, and push the menu button to confirm.
8. Select
Trigger
, and push the menu button to confirm.