Chapter Four: Installation and Operation
© 2018 Cutting Edge Optronics, Inc.
35
Patara I and II Laser User Manual
Menu 1
Menu 2
Menu 3
Parameter
Setting or Value
Comm Setup
Channel 0 Setup
Set Function
Function
Oven TEC
Channel Setup
Comm 0*
Set Temp
See Test Report
Set Tolerance Range
+1.5C
Set output voltage
5.00 V
Set Min Temp
+22.0C
Set Max Temp
+45.0C
Set P Const.
4.0C
Set I Const
2.00 rep/min
Set D Const
0.00 cycles/min
Set Control Function
H-BRIDGE
Set Heat Mult.
1.000
Set Cool Mult
1.000
Sensor Type
TS67
Sensor Offset
0
TEC Enabled
(Note: must be done
last)
Enabled
Interface Setup
Set Trig Out
Mode
Trig out mode
QSW Active High
Interface Setup
Shutter Setup
Shutter FPS
Enabled
Shutter Speed
5 ms
Closed to Standby
Disabled
Interface Setup
Marking Mode
Setup
LM Active
Low
Gate Active
High
FPS Active
High
Utility Functions
Manual Lockout
Disabled
Utility Functions
Humidity Options
Measured Current
Humidity
(reads out
measured humidity)
Adjust Humidity
Threshold
29.5
4.
Verify that there are no objects in the laser beam path except for the negative
lens and power meter.
WARNING.
Wear proper laser safety eyewear to protect the eyes.
5.
Once all the parameters are set correctly, and temperatures of TEC Controller
and chiller are stabilized, set the current at 10 A and press the
EMISSION
button.
6.
Press the
SHUTTER
button to open the laser shutter. Gradually increase the
current until the laser is firing at very low power. Move the negative lens and