User Manual for Patient Monitor
22
4.3.2 Operation Instruction
between the ST segment value and the referenced value. Th
e value is displayed after the measure on “ST
+ 0.000 mV”. The operation is carried out in 4 steps.
First, rotate the Navigation Knob to move the base point (the red cross) horizontally to base line point (the
base line is between the Q wave and the P wave
). At this point, the frozen screen shows “ST+0.xxx mV, Set
Base, Dirc Hor”.
Second, press the Navigation Knob. The screen shows “ST+0.xxx mV, Set Base, Dirc Ver”. Then rotate the
knob to move the base point vertically to the base line point.
Third, press
the Navigation Knob again. The screen shows “ST+0.xxx mV, Set STDot, Dirc Hor”. Rotate the
knob to move the ST point (the yellow cross) horizontally to the point to be measured on the ST segment.
Last, press the Navigation Knob again. The screen
shows “ST+0.xxx mV, Set STDot, Dirc Ver”. Rotate the
knob to move the ST point vertically to the point to be measured on the ST segment.
One the main screen allows pressing the freeze key to enter the ST segment analysis screen.
NOTE: The S point is the end point of S wave, and the T point is the start point of T wave.
4.4 Mode Selection Screen
Press the Navigation Knob in the main screen as shown in Figure 4.2, the operating area shows the mode
selection screen, as shown in Figure 4.12.
Figure 4.12 Trend Menu Screen
In the mode selection screen, rotate the knob to move the gray cursor to the corresponding screen. Press
the knob to enter the screen of SpO
2
, Trend, Recall, ARR, Setup, Color, and File. The following sections from
chapter 4.7 will describe each one of the 8 screens. If you want to exit from this screen, press the Display Key.
ARR