User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
14
-
NIBP monitoring is prohibited to those who have severe hemorrhagic tendency or with
sickle cell disease, or partial bleeding will appear.
Pay attention to the color and sensitivity of the limb when measuring NIBP; make sure
the blood circulation is not blocked. If blocked, the limb will discolor, please stop
measuring or remove the cuff to other positions. Doctor should examine this timely.
Confirm your patient category (adult, pediatric or neonate) before measurement.
Do NOT bind NIBP cuff on limbs with transfusion tube or intubations or skin lesion
area, otherwise, damages may be caused to the limbs.
Prior to use of the cuff, empty the cuff until there is no residual air inside it to ensure
accurate measurement.
Do NOT twist the cuff tube or put heavy things on it.
When unplugging the cuff, hold the head of the connector and pull it out.
3.3.2 SpO
2
Sensor Connection
SpO
2
sensor is a very delicate part. Please follow the steps and procedures in operating it. Failure to
operate it correctly can cause damage to the SpO
2
sensor.
Operation procedure:
1. Connect the SpO2 sensor to the connector labeled “SpO
2
”.
When unplugging the
probe, be sure to hold the head of the connector and pull it out.
2. If the finger clip SpO2 sensor is used, insert one finger into the sensor (index finger,
middle finger or ring finger with short nail length) as shown in the figure below.
Figure 3.4 Finger clip SpO2 sensor placement
3. If the neonate SpO2 sensor is used, please follow Figure 3.5 to connect.