User Manual for Multi-parameter Patient Monitor                                                                          North-vision Tech. Inc. 

 

 

1

  -

 

 

 

Multi-parameter Patient Monitor 

Deluxe-80 

 

 

User Manual 

Version: 1.5

 

 

 

 

 

 

Issued by:    Sara Lo          Date: 5/11/2020 

 
 

Reviewed by: Albert Huang    Date: 5/11/2020 

 
 

 
 
 

Document No: UM-PM8-201507-01

 

North-Vision Tech. Inc.

 

 

 

 

2460

 

Содержание Deluxe-80

Страница 1: ...Patient Monitor North vision Tech Inc 1 Multi parameter Patient Monitor Deluxe 80 User Manual Version 1 5 Issued by Sara Lo Date 5 11 2020 Reviewed by Albert Huang Date 5 11 2020 Document No UM PM8 201507 01 North Vision Tech Inc 2460 ...

Страница 2: ...escribes in accordance with the Vital Signs Monitor s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and storage etc as well as the safety procedures to protect both the user and equipment Refer to the respective chapters for details The Manual is published in English and we have the ultimate righ...

Страница 3: ...tor and remove sensors while during MRI scanning or else induced current could potentially cause burns The monitor may affect the MRI image and the MRI unit may affect the accuracy of monitor measurements Although biocompatibility tests have been performed on all the applied parts some exceptional allergic patients may still have anaphylaxis Do NOT apply to those who suffer from anaphylaxis All th...

Страница 4: ...CTION 13 3 3 1 Blood Pressure Cuff Connection 13 3 3 2 SpO2 Sensor Connection 14 3 3 3 Loading printing paper 15 3 3 4 Battery Installation 16 CHAPTER 4 OPERATIONS 18 4 1 INITIAL MONITORING SCREEN 18 4 1 1 Screen Description 18 4 1 2 Operation Instructions 19 4 2 SPO2 TREND GRAPH DISPLAY 20 4 3 NIBP LIST SCREEN 21 4 3 1 Operation Instructions 21 4 4 SETUP MENU SCREEN 22 4 4 1 SpO2 Setup 22 4 4 2 N...

Страница 5: ...le 32 7 3 2 Points to be noted in SpO2 and Pulse Measuring 32 7 3 3 Clinical Limitations 33 CHAPTER 8 TROUBLESHOOTING 34 8 1 NO DISPLAY ON THE SCREEN 34 8 2 NO BLOOD PRESSURE AND PULSE OXYGEN MEASURES 34 8 3 SYSTEM ALARM 34 CHAPTER 9 MAINTENANCE 35 9 1 SERVICE AND EXAMINATION 35 9 1 1 Daily Examination 35 9 1 2 Routine Maintenance 35 9 1 3 Battery Maintenance 35 9 2 CLEANING STERILIZATION AND DISI...

Страница 6: ...last 12 24 or 96 hours Up to 400 groups of NIBP measurements can be stored and reviewed by list Audible visible alarm with 3 levels of alarm events Nurse call output is available NIBP measurement is applicable to adult pediatric and neonate by patient selection Built in printer is optional to print out waveforms and text information 1 2 Product Name and Model Name North vision Multi parameter pati...

Страница 7: ...User Manual for Multi parameter Patient Monitor North vision Tech Inc 7 1 5 Symbols on the Monitor ...

Страница 8: ... module designed product it consists of NIBP module optional SpO2 module optional main control unit printer module display panel and power supply block etc 1 The SpO2 module detects and calculates pulse rate and oxygen saturation SpO2 and provides plethysmogram and perfusion index as well 2 The NIBP module performs the measurement of blood pressure by non invasive way of oscillometric technology i...

Страница 9: ...ium priority alarm Probe off Green Normal monitoring state 2 SYS display systolic pressure value 3 DIA display diastolic pressure value 4 MAP Display mean arterial pressure or measuring time of the latest group of Module Configuration Instructions According to different needs user can customize the module configuration by choosing the necessary modules This manual shows how to use the monitor of t...

Страница 10: ... pulse intensity 8 LCD panel 9 Patient category indicator for adult for pediatric for neonate Patient category is selected under sub menu Patient Info within the setup menu 10 NIBP start cancel NIBP measurement 11 Alarm silence key Enable disable alarm silence function When the alarm silence indicator on the left of keys is on it means the system is in alarm silence status and it lasts this status...

Страница 11: ...is being recharged ON green OFF the battery is being recharged while this device is off 20 Power button Press power button for 3 seconds to start the monitor or shut off the monitor Note Short time pressing power button for entering the Power Saving Mode screen then according to your need to make the device stay in the power saving mode or exit from power saving mode this function is optional and ...

Страница 12: ...essories for any scratch or deformity especially on connector wire and probe parts If in doubt please contact the local dealer or our company in case of any problems We are to offer you the best solution for your satisfaction 3 2 2 Connecting the Power Supply 1 When powered by AC mains power supply Make sure that the AC mains power supply is within 100 250VAC with 50Hz or 60Hz Use the power cord p...

Страница 13: ... subject Its width should be 2 3 of the length of the upper arm The cuff inflation part should be long enough to permit wrapping 50 80 of the limb concerned See the table below for the dimensions Cuff Type Arm circumference Cuff width Neonate Cuff 6 0cm 9 5cm 3cm Small sizedCuff for Pediatric 6cm 11cm 4 5cm Middle sized Cuff for Pediatric 10cm 19cm 8cm Large sized Cuff for Pediatric 18cm 26cm 10 6...

Страница 14: ...rior to use of the cuff empty the cuff until there is no residual air inside it to ensure accurate measurement Do NOT twist the cuff tube or put heavy things on it When unplugging the cuff hold the head of the connector and pull it out 3 3 2 SpO2 Sensor Connection SpO2 sensor is a very delicate part Please follow the steps and procedures in operating it Failure to operate it correctly can cause da...

Страница 15: ... limb otherwise the NIBP measuring will affect SpO2 measuring and cause the alarm error Using nail polisher or other cosmetic product on the nail may affect the accuracy of measurement The fingernail should be of normal length Do NOT use the damaged SpO2 sensor The SpO2 sensor can not be immerged into water liquor or cleanser completely because the sensor has no capability of waterproofness 3 3 3 ...

Страница 16: ...ckwise and pull the paper out 4 5 steps are the same with the 6 7 steps mentioned above for loading printing paper Figure 3 6 Loading and taking out printing paper 3 3 4 Battery Installation 1 Ensure that the monitor is not connected to AC power supply and the monitor is turned off 2 Open the battery cover and place the battery in the direction as shown in Fig 3 7 to insert the battery into any on...

Страница 17: ...er Patient Monitor North vision Tech Inc 17 Figure 3 7 Battery Installation Note Do not insert battery terminal with its polarities reversed or the monitor can not be started Please take out the battery before transport or storage ...

Страница 18: ...rea see Figure 4 1 The prompt info area contains 3 segment of information status or event indication at the left patient ID number in the middle real time clock at the right also see Figure 4 1 4 1 1 Screen Description Title area PI 3 the perfusion index is 3 it displays only when Setup Menu SpO2 PI Display is set as ON PLETH Mark of SpO2 plethysmogram when PLETH displays in title area the main di...

Страница 19: ...g down timer of NIBP measurement is 120 seconds This prompt message appears only when the NIBP measuring mode is set as AUTO X mute C D 120 the counting down timer of alarm silence is 120 seconds This prompt message appears only when the alarm silence is enabled Patient ID segment 01 Patient ID number Real time clock segment 17 44 59 the current time 4 1 2 Operation Instructions key press this key...

Страница 20: ...24 96 hours if the selecting box stays in trend length option or to move the cursor if the selecting box stays in cursor on option Instructions for viewing the trend curve Select cursor on and press key to confirm and cursor on becomes cursor off then you can press key or key to move the vertical cursor the list box below will display SpO2 value and the time value at the point where the cursor sta...

Страница 21: ...d column is NIBP value and the fourth column is pulse rate measured by NIBP module SYS DIA MAP indicates the value of systolic pressure diastolic pressure mean arterial pressure 4 3 1 Operation Instructions On NIBP List screen if NIBP measurement is more than 5 groups press key or key to scroll up or down through all the measurement values If NIBP measurement is not more than 5 groups the keys or ...

Страница 22: ...There are 7 functional groups for setting parameters SpO2 NIBP Nurse Call System Patient Info Date Time and Default on the Setup Menu Screen 1 Press key or key to shift cursor to corresponding functional group setting 2 Pres key to confirm and enter into corresponding functional parameter setup screen 3 Pres key under the setup menu will print the SpO2 plethysmogram 4 Press to exist from Setup Men...

Страница 23: ... of SpO2 alarm range 0 248 PI display on means PI display is enable off means PI display is disable Operation Instructions 1 Press key or key to move cursor to select parameter 2 Press key to confirm and active this parameter setting 3 Press key or again to adjust or modify parameter value 4 Press key again to confirm and save the setting 5 Press key to return to upper level screen 4 4 2 NIBP Setu...

Страница 24: ...arm range 26 240 mmHg 4 4 3 Nurse Call Figure 4 8 Nurse Call Setup Screen Output level two options low or high output levels are available When the calling system in hospital works in Normal Open mode low level should be selected When the calling system in hospital works in Normal Close mode high level should be selected Source three kinds of alarm sources can trig the nurse call high level alarm ...

Страница 25: ...evel adjustable the set 0 i e no sound key beep to turn on off key beep LANG language selection Two options ENG for English or CHN for Chinese PR SRC source of pulse rate value The default set is SpO2 it can not be adjusted backlite turn on off backlight contract adjust LCD display contract 0 7 level adjustable care mode Demo shows the demo waveforms and data In the demo state all the signals and ...

Страница 26: ...he patient ID is changed all the history data will be cleared that means SpO2 trend graph and NIBP list will become empty 4 4 6 Date Time Figure 4 11 Data Time Setup Screen yy 07 mm 09 dd 21 date setting 07 09 21 shows the date is September 21st 2007 hh 10 mm 15 ss 20 time setting 09 20 21 shows the time is10 15 20 4 4 7 Recover Default Settings On Setup Menu screen press button or button to shift...

Страница 27: ...reen to Power Saving Mode display screen as shown in Figure 4 12 Figure 4 12 Power Saving Mode Press button or button to shift cursor to yes or no and press button to confirm If your selection is yes all the numerical values displayed on digital LEDs display become darker and the monitor stays in power saving mode Short time press power button again to shift screen to Power Saving Mode display scr...

Страница 28: ...mHg Neonate 140 mmHg F Typical measurement range Systolic 20mmHg 290mmHg Diastolic 10mmHg 260mmHg MAP 15mmHg 275mmHg G Measurement accuracy Systolic 5 mmHg Diastolic 5 mmHg MAP 5 mmHg H Measurement mode Manual Auto 5 2 SpO2 Monitoring A Sensor dual wavelength LED B SpO2 measurement range 0 100 C SpO2 measurement accuracy 70 100 2 5 3 Rulse rate Monitoring A Pulse rate measurement range 15bpm 300bp...

Страница 29: ...All Parameters Mode Parameter Alarm setting range Adult Infant Neonate Systolic High 31 250 mmHg 180mmHg 130mmHg 110mmHg Low 30 249 mmHg 60mmHg 50mmHg 50mmHg Diastolic High 21 232 mmHg 120mmHg 90mmHg 90bpm Low 20 231 mmHg 50mmHg 40mmHg 30bpm MAP High 27 242 mmHg 160mmHg 110mmHg 100mmHg Low 26 241 mmHg 50mmHg 40mmHg 30mmHg SpO2 High 1 100 100 100 100 Low 0 90 90 85 85 Pulse Rate High 21 250 bpm 180...

Страница 30: ...artons with foam inside to protect the equipment against damage in the shipping and handling process 6 2 Accessories Item Content Quantity 1 NIBP cuff One set 2 SpO2 probe One set 3 Power cord One set 4 Grounding wire One set 5 Disposable electrode Ten pieces 6 User manual One copy Note The accessories are subject to change Detailed items and quantity see the Packing List ...

Страница 31: ...ood pressure at amplitude of cuff pressure forward reduced according to proper proportion is defined as systolic pressure while the blood pressure at amplitude of cuff pressure backward reduced according to proper proportion is defined as diastolic pressure The maximum change of pulse pressure occurs at these two points They are equivalent to the point with pulse sound and the point without pulse ...

Страница 32: ...heart lung machine 4 Do not take the measurement when the subject uses diuresis or vasodilator 5 When the subject is suffering from major hemorrhage hypovolemic shock and other conditions with rapid blood pressure change or when the subject has too low body temperature the reading will not be reliable for reduced peripheral blood flow will lead to reduced arterial pulsation 6 Subject with hyperadi...

Страница 33: ...ulse substantial pulsating blood stream of subject is required For a subject with weak pulse due to shock low ambient body temperature major bleeding or use of vascular contracting drug the SpO2 waveform PLETH will decrease In this case the measurement will be more sensitive to interference 2 For those with a substantial amount of staining dilution drug such as methylene blue indigo green and acid...

Страница 34: ...g instructions if the cuff leaks and if the inlet is closely connected with the NIBP jack on the side panel Check if the indicator of the pulse oxygen probe flashes and if the pulse oxygen probe is properly connected to the SpO2 jack on the side panel 2 If the problems still exist please contact the dealer 8 3 System Alarm 1 When the parameter value is higher or lower than the alarm limits the ala...

Страница 35: ... yearly maintenance the monitor can be thoroughly inspected by qualified personnel including function and safety examinations If the hospital fails to carry out a satisfactory maintenance program about the monitor it may get disabled and harm the patient s safety and health If there is any indication of cable and transducer damage or they deteriorate they are prohibited from any further use The ad...

Страница 36: ...monitor can be sterilized and disinfected please clean it first Switch off the monitor and disconnect the power cable before cleaning Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage Clean the exterior of the connector only Dilute the cleanser Do not use scrub materials Do not let any liquid flow into the shell or any parts of the monitor Do not let the cl...

Страница 37: ...is not wrapped correctly or is not connected Gas leak Air moving part tube or the cuff leak air Signal weak Very weak signal because of the cuff or the patient has very weak pulse Over extent The measurement range exceeds 255mmHg for neonates over 135 mmHg Over motion The repeated measurement due to moving excessive noise during the stepping inflation and measuring pressure and pulse e g during pa...

Страница 38: ...l No 15 Gongye E 2nd Rd East Dist HsinChu City 30075 Taiwan R O C Tel 886 3 5771038 Fax 886 3 5771039 E mail ahuang north vision com Website http www north vision com Version 1 5 Revised date May 11 2020 Information is subject to change and or updating without notice Copyright 2020 North vision All rights reserved European Representative CMC Medical Devices Drugs S L C Horacio Lengo N 18 CP 29006 ...