Technical Information
Flow Rate:
Typically 0.4 ml/min with 0.9% NaCl solution. The flow rate is very dependent on the drug used.
Aerosol Droplet Size:
Typically 70% of particles generated are less than 5 microns VMD measured using Malvern
Spraytech with 0.9% NaCl solution. The aerosol droplet size is dependent on the drug used.
Medication Cup Volume:
The medication cup will hold 6ml of liquid but is expandable to 24ml using a plug in extension
cup accessory. The controller has an 18 minute timer function therefore if the extension cup is used the
controller will need to be re-started to complete the dose.
Battery Power:
Battery Type: 3.7V Li-ion rechargeable battery. Use the Nortev supplied battery only.
Use of another type of battery may present a risk of fire or explosion.
Operating Time: Approximately 3 x 18 minute treatments from fully charged.
Battery Charging:
Use the Flexineb C1 5V power supply to charge the battery. Alternatively use the supplied USB cable
and plug it into a PC USB port to charge. A fully discharged battery can take 1.5 hours to reach full charge.
Mains Power Supply:
The Flexineb C1 can also be operated using the 5V Power Supply provided with the system.
The same Power Supply charges the battery. Use the Nortev Power Supply only.
Use of an alternative Power Supply may damage the unit or present a risk of fire or explosion.
Environmental Conditions:
Operating: 5°C to 45°C and up to 95% RH
Storage and Transportation: -20°C to +45°C and 15% to 95% RH
Controller Functions
Troubleshooting
Refer to
www.nortev.com
website for detailed information on troubleshooting Flexineb
C1
problems.
Maintenance / Cleaning
Medication Cup:
Cleaning:
It is very important to wash out the Medication Cup after each use to avoid contamination or
damage of the metallic vibrating mesh. Protect the electrical connector from getting wet when cleaning by
using the protective cap provided. To clean out the Medication Cup open up the soft blue Cap and place 1-
2 drops of mild hand dishwashing liquid (
do not use dishwasher detergent
) into the Medication Cup and
fill with distilled water. Close the soft blue Cap and shake the Cup vigorously for about 15 seconds and
squeeze the soft Cap to push water through the mesh whilst shaking. Open the Cap and rinse out from
both sides with distilled water. Shake off excess water and let air dry before re-use.
Disinfection:
To disinfect the outside surfaces of the Medication Cup use anti-bacterial wipes.
CAUTION:
Do not immerse the Medication Cup in water or cleaning chemical solutions!
Do not nebulise or put chemical cleaning solutions inside the Medication Cup as they may
permanently damage the metallic mesh or reduce its life!
Do not use sharp instruments to clean the metallic mesh in the Medication Cup!
The Medication Cup is a replaceable part. For continuity of treatment it is highly recommended to
carry spare Medication Cups!
Mask and Adaptor:
Cleaning:
The Mask and Adaptor can be
washed using clean warm tap water with some dish washing up
liquid. Rinse off with warm clean tap water and dry off before re-use.
Disinfection:
First
disassemble all parts of the Mask and with the Adaptor wash them thoroughly with
clean warm tap water and dish washing up liquid. Rinse the parts off and let them air dry on a clean dry
surface. Disinfect each part by wiping down thoroughly with anti-bacterial wipes
before re-assembly.
Check
www.nortev.com
for updates and video instructions on Cleaning & Disinfecting Flexineb
C1
Controller:
To maintain the function of the electronic controller it must be used and stored in dry conditions. To
maintain the battery function it must be charged when the Operation LED turns Red. Disinfect by wiping down
with anti-bacterial wipes.
CAUTION:
Do not immerse the controller in water or any liquid as this will permanently damage the electronics!
Warranty
Mask, Adaptor, Controller (excluding battery), Cable, Power Supply (charger): 24 months. Battery: 6
months. The Medication Cup is a wearing part and its life is very dependent on drugs nebulised and how
well it is maintained therefore it is not covered by warranty.
Damage caused by improper handling or misuse is not covered by warranty.
Compliance
Flexineb
C1
Small Animal Nebuliser complies with the essential requirements of all relevant European
Directives and Harmonised Standards.
Nortev Limited
,
Unit 18
,
Claregalway Corporate Park
,
Claregalway
,
Co. Galway, Ireland
Operation LED:
Green =
System On
Constant Red=
Low Battery
Warning
Flashing Red
= Battery is dead
and you must re-charge the
battery to continue using the
system.
Flashing Green & Power Off =
Cup not connected, damaged
cup, damaged cable or internal
error.
Contact your customer service
line for advice.
Battery Charging LED:
Constant Red
means
the Flexineb
C1
charger
is connected and the
battery is charging.
Constant Green
means
the battery has reached
full charge.
On / Off
Switch
MS-264 Rev. 01