TDMG0002 Icon-5 Maintenance Guide 

– Version: 02 – Release date: 2020-04-27 

 

 

Icon-5 Maintenance Guide 

Kit description and procedures 

 

This document describes the contents of the maintenance kit and list the corresponding 
maintenance procedures for Norma Icon-5 automated hematology analyzer. 

Special tool required:

 UNF tool 

(not included in the kit) 

Please contact our support team at 

[email protected]

 if further information or 

clarification is required. 

 

Icon-5 Maintenance Kit (HAT827A) 

– for 1-year/10.000 sample period 

Item

 

Part# 

pcs

 

Aperture maintenance kit  

HAT809A 

Peristaltic pump tube set for “gray” and “orange” 

pumps 

HAT800B 

HAT800C 

1 pair of each 

Lubricants (in small containers) 

 

 

-

 

Shell Retinax lubricant (red) 

 

HAT829A 

-

 

Teflon based grease (white)

 

HAT829B 

-

 

Silicone grease (clear)

 

HAT829C 

Piercing needle with UNF screw and “O” ring 

HAT165A 

1  

Greiner BIO Vacutainer NO add w/o rubber in cap 

(White) 

HAT823B 

 

 

Содержание Icon-5

Страница 1: ...ct our support team at support normadiagnostika com if further information or clarification is required Icon 5 Maintenance Kit HAT827A for 1 year 10 000 sample period Item Part pcs Aperture maintenanc...

Страница 2: ...nfection contamination WARNING The analyzer should only be operated without covers with extreme caution moving parts can cause injury High Voltage board HVDCDC produces 180VDC Service marked component...

Страница 3: ...sults in deepening cracks on the tube material which will eventually lead to leaks and reduction or failure in liquid movement The wash head will slowly fail to remove waste from the sampling needle w...

Страница 4: ...s gently pull on the adapter to release it from the groove Pull the tubes off the tube adapters The kit contains two different sets of pump tubes Different tubes must be used for different pumps HAT80...

Страница 5: ...on the tubes to make them sit properly in the pump over the rollers Reinstall the cassette Both ends must CLICK upon insertion Make sure that the cassette is level with the pump body Incorrect alignm...

Страница 6: ...6 TDMG0002 Icon 5 Maintenance Guide Version 02 Release date 2020 04 27 Reinstall the pump Watch out for the external silicone tubes Avoid pinching the tubes and cables between metal plates...

Страница 7: ...e keeps the measuring apertures free of unwanted air and regulates liquid flow during different stages of the measurement process If the rubber valve gets contaminated then it will not be able to clos...

Страница 8: ...rve when the rubber valve base lays flat on the base of its socket Overtightening the screw cap will deform the valve and make its operation impossible Reinstall the aperture module The two pins must...

Страница 9: ...9 TDMG0002 Icon 5 Maintenance Guide Version 02 Release date 2020 04 27 Reinstall the metal shield...

Страница 10: ...aperture A worn silicone sealing ring will allow the aperture to leak and will form contamination around the module and can even cause errors during measurement due to air or liquid leaks Leaks on ape...

Страница 11: ...alves by removing the two screws The ruby is inside the silicone seal Carefully remove it from the aperture half using a tweezer Use a damp cloth immersed into DI water to clean the sides of the apert...

Страница 12: ...by is not symmetrical one side is flat the other side has a spherical dent The spherical side must face the LEFT side aperture housing See picture above There is a cut out in the plastic housing where...

Страница 13: ...ghtening may cause the silicon to flat out fast again and start leaking The aperture halves must have a small even gap between them Do not flatten out the gap between the aperture halves Doing so will...

Страница 14: ...be able to do so use a marker to mark the brake s actual position on two places By doing so it will be easy to find the proper positioning during reassembling Loosen the two small hex screws of the b...

Страница 15: ...cap and can get clogged Use the Norma UNF tool not included to manipulate the sampling needle To access the piercing needle the closed wash head must be pushed away from the needle Use a 2mm hex key t...

Страница 16: ...there is any dirt residuals are on it Use a brush to clean the wash head in and out Use the white grease from the KIT to lubricate the axis the wash head is turning It is possible to put a little drop...

Страница 17: ...tool if needed Overtightening may cause damage to the threads but loose connection will cause leakage in the fluidic system which will show up in erroneous results If there are visible white salt depo...

Страница 18: ...state of all injectors look for salt deposits around their bottom fixing plate In case there is any it is maybe a sign of a slight leakage of that injector or of the related valves Check the state of...

Страница 19: ...ep 11 Reagent sensors and connections Frequency 1 year or 10000 samples part N A Make sure all tubes are properly placed in the reagent sensors It should click in In case there is a liquid detection p...

Страница 20: ...metal tools when replacing the battery you may cause short circuit and a damage the board or the battery Do not touch the surface of the new battery by fingers because contamination salt or moist may...

Страница 21: ...following steps Norma iHClean or similar solution is needed for this process Take the Norma iHClean or hypochlorite solution and immerse the diluent pickup tube into it Run diluent prime 3 times and g...

Страница 22: ...ion if necessary Run the freshly opened normal control sample and check the NEU XY parameter on the results technical screen It should meet with the given NNEU XY values of the control assay sheet In...