11.2 Maintenance by Qualified Individuals
The following activities should only be undertaken by PC support (IT) personnel, equipment
technicians or those with suitable training.
11.3 Companion Software Updates
•
Perform a backup of the data folders to a separate drive as a precaution.
•
Click on the Patch/Update link provided in the email or as given on the
•
Download the Patch/Update file.
•
To install the Patch/Update, click “Run” on the dialog box. No password is required.
Attention
All EMG sensors should be fully charged before a firmware update is performed.
Symptom
:
Problem with the EMG Sensor recognizing the SmartLead
Possible Reason
Remedial Action
SmartLead is disconnected or loose
Check that the SmartLead is inserted Fully
Firmware in the Sensor is outdated. The EMG
Sensor Status LED may be solid Yellow.
Sensor firmware is periodically updated to support
newly designed SmartLeads. Make sure you have
installed the latest version of MR3 and run a
firmware update if needed.
12 Troubleshooting