![Nonin RespSense LS1R-9R Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/nonin/respsense-ls1r-9r/respsense-ls1r-9r_operators-manual_1700102032.webp)
26
7. Verify all parameters display correctly and adjust any alarm limits according to the patient.
8. Verify alarm function/status by simulating alarm situations for all parameters.
9. Visually verify the zero-point of the CO
2
graph is not elevated.
WARNING: If RespSense fails to respond as described, discontinue use and contact
Nonin Technical Service.
CAUTION:
RespSense should only be operated by trained licensed practitioners.
WARNING: Never allow liquids to enter into or to be spilled onto the monitor. If liquid
has penetrated into the monitor it must be checked by Nonin Technical Service.
CAUTION:
Be careful not to drop
RespSense
on the floor or strike it
against hard
surfaces. If such an incident happens, do not use RespSense until a functional test
has been carried out.
CAUTION:
Never open the monitor housing/case. By opening the case you render
your warranty invalid.
!
!
!